2 xplorer pipette, multi-channel, Xplorer pipette, multi-channel – Eppendorf XplorerPlus Adjustment User Manual

Page 19

background image

Error

Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

19

3.1.2

Xplorer pipette, multi-channel

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Error limits ISO

8655

Error

Systematic error

Random error

±

%

±

µL

±

%

±

µL

0.5

-

10

μL

Increment:
0.01

μL

Medium gray
0.1 - 20 μL
40 mm

1

μL

±24.0

±0.24

±16.0

±0.16

5

μL

±4.8

±0.24

±3.2

±0.16

10 μL

±2.4

±0.24

±1.6

±0.16

5

-

100

μL

Increment:
0.01

μL

Yellow
2 - 200

μL
53 mm

10 μL

±16.0

±1.6

±6.0

±0.6

50

μL

±3.2

±1.6

±1.2

±0.6

100

μL

±1.6

±1.6

±0.6

±0.6

15

-

300

μL

Increment:
0.2

μL

Orange
20 – 300 μL
55 mm

30

μL

±26.0

±7.8

±10.0

±3.0

150

μL

±5.2

±7.8

±2.0

±3.0

300

μL

±2.6

±7.8

±1.0

±3.0

50

-

1

200

μL

Increment:
1

μL

Green
50 – 1 250 μL
76 mm

120

μL

±26.67

±32

±10.0

±12.0

600

μL

±5.33

±32

±2.0

±12.0

1

200

μL ±2.66

±32

±1.0

±12.0