2 xplorer pipette, multi-channel, Xplorer pipette, multi-channel – Eppendorf XplorerPlus Adjustment User Manual
Page 19
Error
Eppendorf Xplorer
®
/Xplorer
®
plus
English (EN)
19
3.1.2
Xplorer pipette, multi-channel
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Error limits ISO
8655
Error
Systematic error
Random error
±
%
±
µL
±
%
±
µL
0.5
-
10
μL
Increment:
0.01
μL
Medium gray
0.1 - 20 μL
40 mm
1
μL
±24.0
±0.24
±16.0
±0.16
5
μL
±4.8
±0.24
±3.2
±0.16
10 μL
±2.4
±0.24
±1.6
±0.16
5
-
100
μL
Increment:
0.01
μL
Yellow
2 - 200
μL
53 mm
10 μL
±16.0
±1.6
±6.0
±0.6
50
μL
±3.2
±1.6
±1.2
±0.6
100
μL
±1.6
±1.6
±0.6
±0.6
15
-
300
μL
Increment:
0.2
μL
Orange
20 – 300 μL
55 mm
30
μL
±26.0
±7.8
±10.0
±3.0
150
μL
±5.2
±7.8
±2.0
±3.0
300
μL
±2.6
±7.8
±1.0
±3.0
50
-
1
200
μL
Increment:
1
μL
Green
50 – 1 250 μL
76 mm
120
μL
±26.67
±32
±10.0
±12.0
600
μL
±5.33
±32
±2.0
±12.0
1
200
μL ±2.66
±32
±1.0
±12.0
This manual is related to the following products: