1 1-point adjust, Point adjust – Eppendorf XplorerPlus Adjustment User Manual
Page 12
Change adjustment
Eppendorf Xplorer
®
/Xplorer
®
plus
English (EN)
12
2.7.1
1-point adjust.
After you enter the density, the selected volume and the corresponding mean value of the weighing results,
the Xplorer pipette will determine a correction factor. The factor is only correct for the selected volume and
the selected work technique. However, the factor is used for the entire volume range of the Xplorer pipette.
You must carry out a gravimetric test to check the validity of the factor.
Example of the
1-point adjust.
• The systematic error of a 40% salt solution should be improved by a 1-point adjustment for the
pipetting of 500
μL.
• You use a 1
000
μL Xplorer pipette with the volume range 50
–
1
000
μL
with
1
000
μL
epT.I.P.S.
pipette
tips.
• The adjustment
Factory settings
is active for the Xplorer pipette.
• The density of the salt solution is, for example, 1.43
mg/μL (mg/μL = g/mL) at 20°C.
• In preliminary tests you have optimized your work technique (speed, handling blow-out (blow), angle of
inclination and working speed) so that you have achieved the best possible precision (random error) and
handling for your task.
• You have carried out at least 10 weighings for the 500
μL on a fine balance.
• The calculated mean value of the weighings is, for example, 709
mg.
Inputting the data
1. Move the selection dial to
Opt.
2. Mark the
Adjustment
option with the rocker. Press the
Select
softkey.
3. Mark the adjustment
1-point adjust.
with the rocker. Press the
Select
softkey.
4. Set the density with the rocker (example: 1.43
mg/μL). Press the
Next
softkey.
5. Set the first volume with the rocker (example: 500
μL). Press the
Next
softkey.
6. Set the mean value of the weighings with the rocker (example: 709
mg). Press the
Next
softkey.
For the 5
mL and 10
mL Xplorer pipettes enter the values in mL and g.
7. Confirm the confirmation prompt with the
Yes
softkey.
8. To check this, turn the selection dial to a dispensing mode.
If the adjustment
1-point adjust.
was successfully activated, the header of the display will show the symbols:
With a density of 1.43
mg/μL,
709
mg is equivalent to a volume of 495.8
μL.
Without
changing
the adjustment, pipetting 500
μL should result in a lesser finding of 4.2
μL
(approx.
0.8
%).