beautypg.com

1 1-point adjust, Point adjust – Eppendorf XplorerPlus Adjustment User Manual

Page 12

background image

Change adjustment
Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

12

2.7.1

1-point adjust.

After you enter the density, the selected volume and the corresponding mean value of the weighing results,
the Xplorer pipette will determine a correction factor. The factor is only correct for the selected volume and
the selected work technique. However, the factor is used for the entire volume range of the Xplorer pipette.
You must carry out a gravimetric test to check the validity of the factor.

Example of the

1-point adjust.

• The systematic error of a 40% salt solution should be improved by a 1-point adjustment for the

pipetting of 500

μL.

• You use a 1

000

μL Xplorer pipette with the volume range 50

1

000

μL

with

1

000

μL

epT.I.P.S.

pipette

tips.

• The adjustment

Factory settings

is active for the Xplorer pipette.

• The density of the salt solution is, for example, 1.43

mg/μL (mg/μL = g/mL) at 20°C.

• In preliminary tests you have optimized your work technique (speed, handling blow-out (blow), angle of

inclination and working speed) so that you have achieved the best possible precision (random error) and
handling for your task.

• You have carried out at least 10 weighings for the 500

μL on a fine balance.

• The calculated mean value of the weighings is, for example, 709

mg.

Inputting the data
1. Move the selection dial to

Opt.

2. Mark the

Adjustment

option with the rocker. Press the

Select

softkey.

3. Mark the adjustment

1-point adjust.

with the rocker. Press the

Select

softkey.

4. Set the density with the rocker (example: 1.43

mg/μL). Press the

Next

softkey.

5. Set the first volume with the rocker (example: 500

μL). Press the

Next

softkey.

6. Set the mean value of the weighings with the rocker (example: 709

mg). Press the

Next

softkey.

For the 5

mL and 10

mL Xplorer pipettes enter the values in mL and g.

7. Confirm the confirmation prompt with the

Yes

softkey.

8. To check this, turn the selection dial to a dispensing mode.

If the adjustment

1-point adjust.

was successfully activated, the header of the display will show the symbols:

With a density of 1.43

mg/μL,

709

mg is equivalent to a volume of 495.8

μL.

Without

changing

the adjustment, pipetting 500

μL should result in a lesser finding of 4.2

μL

(approx.

0.8

%).

This manual is related to the following products: