5 volume limit, 6 additional volumes during aspiration, Volume limit – Eppendorf XplorerPlus Adjustment User Manual

Page 23: Additional volumes during aspiration

background image

Volume limit

Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

23

5

Volume limit

6

Additional volumes during aspiration

ep

Dualfilter

T.I.P.S.

Volume

limit

Ads/Dis

Reverse pipetting

10 μL

Not required

Not required

100

μL

Not required

Not required

200

μL

Not required

Not required

300 μL

270

μL

260

μL

1,000

μL

Not

required

950

μL

1

250

μL Long*

Not required

1

130

μL

5

mL

4.9

mL

4.6

mL*

10

mL*

9.75

mL

8.5

mL*

* When using these filter tips we recommend to use the adjustment

epTIPS long

in order to increase the accuracy of dispensing. The

volume limit stated applies only to the changed adjustment.

** We also recommend these restrictions on volume limits for "reverse
pipetting" when using the 5

mL and 10

mL epT.I.P.S. without filter.

Xplorer
Single-channel and
multi-channel

Additional volume during aspiration

Ads/Dis

Reverse pipetting

10 μL

Approx. 0.6 μL

Approx. 2 μL

100

μL

Approx. 7 μL

Approx. 20 μL

300 μL

Approx. 20 μL

Approx. 60 μL

1,000

μL

Approx. 65 μL

Approx. 200 μL

1

200

μL

Approx. 140 μL

Approx. 220 μL

5

mL

Approx. 0.3 mL

Approx. 1 mL

10

mL

Approx. 0.55 mL

Approx. 1.8 mL

This manual is related to the following products: