3 error, 1 xplorer pipette, single-channel, Error 3.1 – Eppendorf XplorerPlus Adjustment User Manual
Page 18: Xplorer pipette, single-channel

Error
Eppendorf Xplorer
®
/Xplorer
®
plus
English (EN)
18
3
Error
3.1
Limiting values for measurement deviations according to ISO 8655 ddd
3.1.1
Xplorer pipette, single-channel
Model
Test tip
epT.I.P.S. Color code
Volume range
Length
Testing
volume
Error limits ISO
8655
Error
Systematic error
Random error
±
%
±
µL
±
%
±
µL
0.5
-
10
μL
Increment:
0.01
μL
Medium gray
0.1 - 20 μL
40 mm
1
μL
±12.0
±0.12
±8.0
±0.08
5
μL
±2.4
±0.12
±1.6
±0.08
10 μL
±1.2
±0.12
±0.8
±0.08
5
-
100
μL
Increment:
0.01
μL
Yellow
2 - 200
μL
53 mm
10 μL
±8.0
±0.8
±3.0
±0.3
50
μL
±2.0
±0.8
±0.6
±0.3
100
μL
±0.8
±0.8
±0.3
±0.3
15
-
300
μL
Increment:
0.2
μL
Orange
20 – 300 μL
55 mm
30
μL
±13.0
±3.9
±5.0
±1.5
150
μL
±3.0
±3.9
±1.0
±1.5
300
μL
±1.3
±3.9
±0.5
±1.5
50 - 1
000
μL
Increment:
1
μL
Blue
50 - 1
000
μL
71 mm
100
μL
±8.0
±8.0
±3.0
±3.0
500
μL
±2.0
±8.0
±0.6
±3.0
1,000
μL ±0.8
±8.0
±0.3
±3.0
0.25
-
5
mL
Increment:
5
μL
Purple
0.1 – 5 mL
120 mm
0.5
mL
±8.0
±40.0
±3.0
±15.0
2.5
mL
±1.6
±40.0
±0.6
±15.0
5.0
mL
±0.8
±40.0
±0.3
±15.0
0.5
-
10
mL
Increment:
10 μL
Turquoise
1 – 10 mL
165 mm
1.0
mL
±6.0
±60.0
±3.0
±30.0
5.0
mL
±1.2
±60.0
±0.6
±30.0
10.0
mL
±0.6
±60.0
±0.3
±30.0
You can select a volume up to 0.20
mL
for
the
5
mL Xplorer pipette. The handling significantly
affects the achievable errors for the 0.2
mL-0.5
mL volume range.