beautypg.com

3 error, 1 xplorer pipette, single-channel, Error 3.1 – Eppendorf XplorerPlus Adjustment User Manual

Page 18: Xplorer pipette, single-channel

background image

Error
Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

18

3

Error

3.1

Limiting values for measurement deviations according to ISO 8655 ddd

3.1.1

Xplorer pipette, single-channel

Model

Test tip
epT.I.P.S. Color code
Volume range
Length

Testing
volume

Error limits ISO

8655

Error

Systematic error

Random error

±

%

±

µL

±

%

±

µL

0.5

-

10

μL

Increment:
0.01

μL

Medium gray
0.1 - 20 μL
40 mm

1

μL

±12.0

±0.12

±8.0

±0.08

5

μL

±2.4

±0.12

±1.6

±0.08

10 μL

±1.2

±0.12

±0.8

±0.08

5

-

100

μL

Increment:
0.01

μL

Yellow
2 - 200

μL
53 mm

10 μL

±8.0

±0.8

±3.0

±0.3

50

μL

±2.0

±0.8

±0.6

±0.3

100

μL

±0.8

±0.8

±0.3

±0.3

15

-

300

μL

Increment:
0.2

μL

Orange
20 – 300 μL
55 mm

30

μL

±13.0

±3.9

±5.0

±1.5

150

μL

±3.0

±3.9

±1.0

±1.5

300

μL

±1.3

±3.9

±0.5

±1.5

50 - 1

000

μL

Increment:
1

μL

Blue
50 - 1

000

μL

71 mm

100

μL

±8.0

±8.0

±3.0

±3.0

500

μL

±2.0

±8.0

±0.6

±3.0

1,000

μL ±0.8

±8.0

±0.3

±3.0

0.25

-

5

mL

Increment:
5

μL

Purple
0.1 – 5 mL
120 mm

0.5

mL

±8.0

±40.0

±3.0

±15.0

2.5

mL

±1.6

±40.0

±0.6

±15.0

5.0

mL

±0.8

±40.0

±0.3

±15.0

0.5

-

10

mL

Increment:
10 μL

Turquoise
1 – 10 mL
165 mm

1.0

mL

±6.0

±60.0

±3.0

±30.0

5.0

mL

±1.2

±60.0

±0.6

±30.0

10.0

mL

±0.6

±60.0

±0.3

±30.0

You can select a volume up to 0.20

mL

for

the

5

mL Xplorer pipette. The handling significantly

affects the achievable errors for the 0.2

mL-0.5

mL volume range.

This manual is related to the following products: