1 pswrd. protection, 2 factory settings, Pswrd. protection – Eppendorf XplorerPlus Adjustment User Manual

Page 8: Factory settings

background image

Change adjustment
Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

8

2.1

Pswrd. protection

For the Xplorer plus, the adjustment can be protected with a four-digit numeric password. The password
for the

Edit

mode and

Adjustment

option can vary. The password can be changed or deactivated after it has

been entered. Please contact Eppendorf

AG Application Support or conduct a reset if you lose the

password.

Press the rocker upward to activate the password protection.

ON

appears in the display.

Press the

Select

softkey.

Use the rocker to select the digit.

Press the

Next

softkey to switch to the next field.

Enter data in all of the fields.

Press the

Save

softkey to save the password.

Reenter the password after completing the saving procedure. Press the

Enter

softkey.

Password protection is now active.

2.2

Factory settings

To aspirate liquid into a pipette tip, a piston is moved upwards in a cylinder in the Xplorer pipette. The
pipette tip is approximately conical in shape. This requires that the piston stroke in the cylinder be adapted
for aspiration in the conical pipette tip. The mathematical correction of the piston stroke is different for
each volume size of the Xplorer pipette. The respective correction of the piston stroke has been optimized
in the factory setting for the test tip specified in the technical data.

Only available for the Xplorer plus pipette

The adjustment can no longer be changed if the password has been lost.

The factory setting is usually the best setting for all dispensing tasks with purely aqueous
solutions at room temperature.
When verifying the measurement deviations listed in the technical data, the factory setting
must be used with the test tip listed in the technical data.