beautypg.com

Eppendorf XplorerPlus Adjustment User Manual

Page 13

background image

Change adjustment

Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

13

Abb. 2-1: Example: Correction of the piston stroke with "

Abb. 2-2: 1-point adjust.

Abb. 2-3: "

Fig. 2-3:

Example: Correction of the piston stroke with "

1-point adjust.

"

In this example, the piston stroke of the factory setting is increased by a factor. Strictly speaking, the
correction only applies to the test volume (example: 500

μL). However, the correction with this factor is

used for the entire volume range. The correction is different to that for a mechanical pipette. A correction
with a factor for a small volume results in a smaller stroke correction than with a large volume (see Fig. 2-1
on p. 13).

For comparison purposes: The stroke of a mechanical pipette can only be changed by a fixed volume (see
Fig. 2-4 on p. 14). This volume change applies to the entir
e measurement range of the mechanical pipette.
The existing adjustment is changed in parallel by a fixed amount. If, for example, the volume was increased
by 10

μL in order to reach the nominal volume of a 1

000

μL

Research

plus,

this

increase

of

10

μL will also

apply to all volumes below. The correction of the piston stroke of the Xplorer

pipette with a factor

corresponds rather to practical requirements.

What happens in the Xplorer pipette?

The Xplorer pipette uses the data entered to carry out the volume calculation and the correction of the
piston stroke.

Example values for the calculation of the
correction factor

• Mean value of the weighings: 709

mg

• Density of the liquid: 1.43

mg/μL

• Nominal volume: 500

μL

Calculation of the actual volume:
709

mg ÷ 1.43

mg/μL = 495.8

μL

Calculation of the volume difference:
500

μL – 495.8

μL

=

4.2

µL

The piston stroke of the adjustment

Factory settings

is increased by a factor. The factor is formed from the

specified data.

min.

Factory setting

Legend:

1-point adjustment

max.

Piston stroke
of Xplorer

max.

Volume
display of Xplorer

25 %

50 %

75 %

This manual is related to the following products: