1 xplorer pipette, single-channel, Xplorer pipette, single-channel – Eppendorf XplorerPlus Adjustment User Manual
Page 20

Error
Eppendorf Xplorer
®
/Xplorer
®
plus
English (EN)
20
3.2
Eppendorf AG limiting values for measurement deviations
3.2.1
Xplorer pipette, single-channel
pipette
Increment
Pipette tip
Color code
Volume range
Length
Testing
volume
Error limits
Error
Systematic error
Random error
µL
±µL
±%
±µL
±%
±µL
0.5
μL – 10
μL
Increment:
0.01
μL
Medium gray
0.1
μL – 20
μL
40 mm
1
μL
±2.5
±0.025
±1.8
±0.018
5
μL
±1.5
±0.075
±0.8
±0.04
10 μL
±1.0
±0.1
±0.4
±0.04
5
μL–
100
μL
Increment:
0.1
μL
Yellow
2
μL
- 200
μL
53
mm
10 μL
±2.0
±0.2
±1.0
±0.1
50
μL
±1.0
±0.5
±0.3
±0.15
100
μL
±0.8
±0.8
±0.2
±0.2
15
μL – 300
μL
Increment:
0.2
μL
Orange
15
μL – 300 μL
55
mm
30
μL
±2.5
±0.75
±0.7
±0.21
150
μL
±1.0
±1.5
±0.3
±0.45
300
μL
±0.6
±1.8
±0.2
±0.6
50
μL – 1
000
μL
Increment:
1
μL
Blue
50
μL – 1000
μL
71 mm
100
μL
±3.0
±3.0
±0.6
±0.6
500
μL
±1.0
±5.0
±0.2
±1.0
1000
μL
±0.6
±6.0
±0.2
±2.0
mL
mL
mL
±%
±mL
±%
±mL
0.25
mL
–
5
mL
Increment:
0.005
mL
Purple
0.1
mL – 5
mL
120 mm
0.5
mL
±3.0
±15.0
±0.6
±3.0
2.5
mL
±1.2
±30.0
±0.25
±6.25
5
mL
±0.6
±30.0
±0.15
±7.5
0.5
mL – 10
mL
Increment:
0.01
mL
Turquoise
1
mL – 10
mL
165 mm
1
mL
±3.0
±30.0
±0.60
±6.0
5
mL
±0.8
±40.0
±0.20
±10.0
10
mL
±0.6
±60.0
±0.15
±15.0