beautypg.com

1 xplorer pipette, single-channel, Xplorer pipette, single-channel – Eppendorf XplorerPlus Adjustment User Manual

Page 20

background image

Error
Eppendorf Xplorer

®

/Xplorer

®

plus

English (EN)

20

3.2

Eppendorf AG limiting values for measurement deviations

3.2.1

Xplorer pipette, single-channel

pipette
Increment

Pipette tip
Color code
Volume range
Length

Testing
volume

Error limits

Error

Systematic error

Random error

µL

±µL

±%

±µL

±%

±µL

0.5

μL – 10

μL

Increment:
0.01

μL

Medium gray
0.1

μL – 20

μL

40 mm

1

μL

±2.5

±0.025

±1.8

±0.018

5

μL

±1.5

±0.075

±0.8

±0.04

10 μL

±1.0

±0.1

±0.4

±0.04

5

μL–

100

μL

Increment:
0.1

μL

Yellow
2

μL

- 200

μL

53

mm

10 μL

±2.0

±0.2

±1.0

±0.1

50

μL

±1.0

±0.5

±0.3

±0.15

100

μL

±0.8

±0.8

±0.2

±0.2

15

μL – 300

μL

Increment:
0.2

μL

Orange
15

μL – 300 μL

55

mm

30

μL

±2.5

±0.75

±0.7

±0.21

150

μL

±1.0

±1.5

±0.3

±0.45

300

μL

±0.6

±1.8

±0.2

±0.6

50

μL – 1

000

μL

Increment:
1

μL

Blue
50

μL – 1000

μL

71 mm

100

μL

±3.0

±3.0

±0.6

±0.6

500

μL

±1.0

±5.0

±0.2

±1.0

1000

μL

±0.6

±6.0

±0.2

±2.0

mL

mL

mL

±%

±mL

±%

±mL

0.25

mL

5

mL

Increment:
0.005

mL

Purple
0.1

mL – 5

mL

120 mm

0.5

mL

±3.0

±15.0

±0.6

±3.0

2.5

mL

±1.2

±30.0

±0.25

±6.25

5

mL

±0.6

±30.0

±0.15

±7.5

0.5

mL – 10

mL

Increment:
0.01

mL

Turquoise
1

mL – 10

mL

165 mm

1

mL

±3.0

±30.0

±0.60

±6.0

5

mL

±0.8

±40.0

±0.20

±10.0

10

mL

±0.6

±60.0

±0.15

±15.0

This manual is related to the following products: