R (according to din 58 970) – Eppendorf Centrifuge 5427 R User Manual
Page 49
![background image](/manuals/551153/49/background.png)
Technical data
Centrifuge 5427 R
English (EN)
49
9.5
Acceleration times and deceleration times for the Centrifuge 5427
R
(according to DIN 58 970)
Rotor
Acceleration time/Deceleration time
Mains voltage
230
V
120
V
100
V
FA-45-12-17
Without soft ramp Acceleration time
27
s
29
s
29
s
Deceleration time
24
s
25
s
25
s
With soft ramp
Acceleration time
39
s
39
s
39
s
Deceleration time
39
s
39
s
39
s
FA-45-24-11
Without soft ramp Acceleration time
18
s
19
s
19
s
Deceleration time
18
s
19
s
19
s
With soft ramp
Acceleration time
29
s
29
s
29
s
Deceleration time
31
s
31
s
31
s
FA-45-24-11-Kit
Without soft ramp Acceleration time
21
s
22
s
22
s
Deceleration time
21
s
21
s
21
s
With soft ramp
Acceleration time
32
s
32
s
32
s
Deceleration time
31
s
31
s
31
s
FA-45-30-11
Without soft ramp Acceleration time
21
s
22
s
22
s
Deceleration time
18
s
19
s
19
s
With soft ramp
Acceleration time
32
s
32
s
32
s
Deceleration time
33
s
33
s
33
s
F-45-30-11
Without soft ramp Acceleration time
21
s
22
s
22
s
Deceleration time
18
s
19
s
19
s
With soft ramp
Acceleration time
29
s
29
s
31
s
Deceleration time
32
s
32
s
32
s
FA-45-48-11
Without soft ramp Acceleration time
28
s
29
s
29
s
Deceleration time
22
s
23
s
23
s
With soft ramp
Acceleration time
36
s
36
s
39
s
Deceleration time
35
s
35
s
35
s
F-45-48-11
Without soft ramp Acceleration time
28
s
29
s
29
s
Deceleration time
22
s
23
s
23
s
With soft ramp
Acceleration time
36
s
36
s
36
s
Deceleration time
35
s
35
s
35
s
F-45-48-PCR
Without soft ramp Acceleration time
11
s
12
s
12
s
Deceleration time
12
s
13
s
13
s
With soft ramp
Acceleration time
22
s
22
s
22
s
Deceleration time
22
s
22
s
22
s
S-24-11-AT
Without soft ramp Acceleration time
18
s
18
s
18
s
Deceleration time
17
s
17
s
17
s
With soft ramp
Acceleration time
29
s
29
s
29
s
Deceleration time
30
s
30
s
30
s