2 error messages, Error messages – Eppendorf Centrifuge 5427 R User Manual
Page 42
Troubleshooting
Centrifuge 5427 R
English (EN)
42
7.2
Error messages
Proceed as follows if the following error messages appear:
1. Remove fault (see remedy).
2. If necessary, repeat centrifugation.
Symptom/message Cause
Remedy
ERR 1
Rotor not detected.
Check rotor.
If this error message appears again,
test with a different rotor.
Err 2
Electronics fault.
Switch centrifuge off and back on
again after > 20
s.
NO RPM (ERR 3)
Error in speed measuring system.
Leave the device switched on until the
error message disappears (10 s or
6
min).
Err 5
Prohibited opening of lid or lid switch is
defective during a run.
Wait for rotor to stop.
Err 6
• Error in the drive electronics.
• Drive overheated.
Repeat the run.
If this error message appears again,
switch centrifuge off and back on
again after > 20
s.
Allow the drive to cool down for at
least 15
min.
Err 7
Major deviation in the speed control.
1. Wait for rotor to stop.
2. Tighten the rotor.
Err 8
Drive fault.
1. Wait for rotor to stop.
2. Repeat the run.
ERR 9
to
ERR 14
Electronics error.
Switch centrifuge off and back on
again after > 20
s.
IMBAL (ERR 15)
Rotor is asymmetrically loaded.
Load the rotor symmetrically and
balance.
ERR 16
to
ERR 17
Electronics error.
Switch centrifuge off and back on
again after > 20
s.
Err 18
Temperature deviation from set value in
the rotor chamber: ±5 °C.
Check the settings.
Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.
Err 19
Condenser overheated.
Check unhindered air circulation
through the air slots.
Allow device to cool down.
Err 20
Temperature sensor in rotor chamber is
faulty.
Switch centrifuge off and back on
again after > 20
s.
Err 21
Temperature sensor on condenser is
faulty.
Switch centrifuge off and back on
again after > 20 s.