5 continuous cooling, 6 endless operation of continuous cooling, Continuous cooling – Eppendorf Centrifuge 5427 R User Manual
Page 32: Endless operation of continuous cooling
Operation
Centrifuge 5427 R
English (EN)
32
5.5.5
Continuous cooling
Continuous cooling
Continuous cooling maintains the rotor chamber at the set temperature if the rotor stops.
• During continuous cooling the display shows the set temperature.
• To prevent the rotor chamber from freezing or condensation from forming, the temperature does not go
below 4
°C
, irrespective of the set temperature.
• If the rotor stops, temperature control is slower than during centrifugation or a temperature control run.
ECO shut-off
Continuous cooling is switched off if the centrifuge is not used for longer than the preset time. The
centrifuge switches to standby mode.
• Default setting: Continuous cooling ends after 8
h.
• Continuous cooling can be limited to 1
h,
2
h or 4
h.
• ECO shut-off can be switched off (continuous cooling set to endless operation).
Limit continuous cooling to 1
h
(2
h,
4
h).
1. Press the menu/enter key.
2. Select
TEMP
using the arrow key. Confirm with the menu/enter key.
3. Select
1
h
(
2
h,
4
h
) using the arrow key. Confirm with the menu/enter key.
4. To activate ECO shut-off after 1
h,
2
h
or
4
h, open and close again the centrifuge lid after the end of the
run.
The device switches to standby mode after 1
h,
2
h or 4
h. The display shows
EP
.
5.5.6
Endless operation of continuous cooling
The ECO shut-off function can be switched off. Continuous cooling is changed to endless operation.
• Endless operation can shorten the service life of the compressor.
• The rotor chamber may freeze.
1. Press the menu/enter key.
2. Select
TEMP
using the arrow key. Confirm with the menu/enter key.
3. Select
oo
using the arrow key. Confirm with the menu/enter key.
Ending continuous cooling
4. Open the centrifuge lid to end continuous cooling.
• If the centrifuge lid is not opened and closed again after the run, ECO shut-off engages
after 8
h (default setting).
• Open the centrifuge lid to end continuous cooling early.