6 rotor f-45-48-5-pcr, 7 rotor s-24-11-at, Rotor f-45-48-5-pcr – Eppendorf Centrifuge 5427 R User Manual
Page 14: Rotor s-24-11-at

Product description
Centrifuge 5427 R
English (EN)
14
2.4.6
Rotor F-45-48-5-PCR
2.4.7
Rotor S-24-11-AT
This rotor is only intended for use with 1.5
mL/2.0
mL
tubes.
The following tubes must not be used in this rotor:
• Adapters for 0.2 mL, 0.4 mL, 0.5 mL and 0.6 mL tubes and the corresponding tubes.
• Spin columns
Rotor F-45-48-5-PCR
Fixed-angle rotor for 48 PCR
tubes
Max. g-force:
11
710
×
g
Max. speed:
10
500
rpm
Max. load (tube and
contents):
48
×
0.43
g
Tubes
Tube
Adapter
Adapter base
Max. g-force:
Capacity
Tube diameter
Max. speed
Tubes per adapter/
rotor
Centrifugation radius
–
conical
11
710
×
g
0.2
mL
Ø
6
mm
10
500
rpm
–/48
9.5
cm
PCR strips
–
conical
11
710
×
g
0.2
mL
Ø
6
mm
10
500
rpm
–/6
×
8
9.5
cm
Rotor S-24-11-AT
Bucket for tubes:
4
×
1.5
mL/2.0
mL
Max. g-force:
16
049
×
g
Max. speed:
12
700
rpm
Max. load per bucket (tube
and contents):
4
×
3.75
g
Vessels
Vessel
Adapters
Adapter base
Max. g-force:
Capacity
Order no.
(international)
Tube diameter
Max. speed
Vessels per
adapter/rotor
Centrifugation radius
Vessel
–
–
16
049
×
g
1.5
mL/2.0
mL
Ø
11
mm
12
700
rpm
-/24
8.9
cm