Maintenance
Centrifuge 5427 R
English (EN)

40

How to proceed in case of glass breakage

1. Remove all splinters and glass powder from the rotor chamber and accessories.

2. Thoroughly clean the rotor and rotor chamber. Thoroughly clean the bores of the fixed-angle rotors, in

particular.

3. If required, replace the adapters to prevent any further damage.

4. Regularly check the rotor bores for deposits and damage.

6.5

Replacing fuses

The fuse holder is located between the mains connection socket and the mains power switch.

1. Disconnect the power plug.

2. Press the upper and lower end of the plastic springs 1 together and pull the fuse holder 2 fully out.

3. Replace faulty fuses and reinsert the fuse holder. Make sure that the guiding rail 3 is positioned

correctly.

6.6

Decontamination before shipment

If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer
for disposal please note the following:

DANGER! Electric shock.

Switch off the device and disconnect the power plug before starting maintenance or
cleaning work.

WARNING! Risk to health from contaminated device

1. Follow the instructions in the decontamination certificate. It is available as a PDF file on

our webpage (www.eppendorf.com/decontamination).

2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.