Eppendorf Centrifuge 5427 R User Manual
Page 43
Troubleshooting
Centrifuge 5427 R
English (EN)
43
Err 22
Electronics fault.
Switch centrifuge off and back on
again after > 20
s.
NO E-FAN (ERR 23)
Electronic fan is faulty.
Switch centrifuge off and back on
again after > 20
s.
Err 24
Failure at the compressor.
Allow the centrifuge to cool down and
repeat the run.
INT (ERR 25)
Power failure during a run.
Check the mains connection.
Err 28
Error during speed check.
Switch centrifuge off and back on
again after > 20
s.
Err 29
Set g-force/speed too high, e.g. after a
rotor change (see Automatic rotor
detection on p. 27).
Check the g-force/speed.
Repeat the run.
LID (ERR 30)
• Centrifuge lid cannot be locked.
• Centrifuge lid cannot be released.
Try again to close centrifuge lid.
Switch the centrifuge off and back on.
Press the open key.
If the error occurs repeatedly:
Switch off centrifuge.
Actuate emergency release(see
Emergency release on p. 44).
OFF
• Displayed after power off.
Symptom/message Cause
Remedy
- epMotion 96 (76 pages)
- epMotion 5070 (100 pages)
- epMotion 5075 (130 pages)
- Centrifuge 5427 R (104 pages)
- White Paper 14 (8 pages)
- Rolling Cabinet (34 pages)
- Mastercycler nexus (118 pages)
- Mastercycler nexus (142 pages)
- Concentrator plus (New Design) (48 pages)
- Concentrator plus (43 pages)
- Easypet 3 (38 pages)
- Xplorer (74 pages)
- Xplorer Adjustment (26 pages)
- AF2200 Plate Reader (78 pages)
- AF2200 Plate Reader (72 pages)
- G0.5 µPlate (32 pages)
- BioSpectrometer basic (104 pages)
- BioSpectrometer kinetic (106 pages)
- BioSpectrometer fluorescence (102 pages)
- Micro Test Tubes (5 pages)
- Microplates (10 pages)
- PiezoXpert (34 pages)
- Eporator (38 pages)
- MiniSpin (20 pages)
- MiniSpin (25 pages)
- Centrifuge 5702 (32 pages)
- 5702 Centrifuge (27 pages)
- 5702 Centrifuge (32 pages)
- C5702 RH Centrifuge (32 pages)
- 5418 Centrifuge (48 pages)
- 5418 Centrifuge (80 pages)
- 5424 Centrifuge (71 pages)
- 5424 Centrifuge (44 pages)
- 5430 Centrifuge (88 pages)
- 5430 Centrifuge (130 pages)
- 5804 Centrifuge (127 pages)
- 5804 Centrifuge (129 pages)
- 5804 Centrifuge (95 pages)
- TransferMan4 r (102 pages)
- TransferMan4 m (96 pages)
- InjectMan 4 (100 pages)
- InjectMan NI 2 (60 pages)
- InjectMan NI 2 (16 pages)
- PatchMan NP 2 (53 pages)