beautypg.com

Eppendorf Centrifuge 5427 R User Manual

Page 26

background image

Operation
Centrifuge 5427 R
English (EN)

26

The

Back

menu item can also be found in both menu levels.

Back

in menu level 2: return to menu level 1.

Back

in menu level 1: exit menu.

SHORT

Set the speed of the short spin
centrifugation.
No SOFT function with short spin
centrifugation.

MAX

Short spin centrifugation at the maximum
speed of the inserted rotor

Set

Short spin centrifugation with set speed
(g-force

(rcf) or speed

(rpm)).

Temp

Set the time limit for continuous cooling
("ECO shut-off") (see p. 30).

8 h

Default setting: Continuous cooling ends
after 8

h.

1 h
2

h
4

h

To limit continuous cooling after a run to
1

h,

2

h, or 4

h, the centrifuge lid must be

opened and closed again after the run.

oo

Endless operation of continuous cooling.

Alarm

Switch the loudspeaker on or off.

On

Switch on loudspeaker.

OFF

Switch off loudspeaker.

VOL

Set the volume.

VOL1

VOL5

The volume of the loudspeaker can be set
to 5 levels (

VOL1

to

VOL5

). The

loudspeaker must be switched on for the
adjustment to be audible.

SLEEP

Switch the standby mode on or off.
If the centrifuge has not been used for
15

min, it will switch to standby.

EP

then

appears in the display.
Press a key or close the centrifuge lid to
exit the standby mode.

On

Standby mode activated.

OFF

Standby mode deactivated.

Menu level 1 (M 1)

Menu level 2 (M 2)

Display