MCA Query Commands:

Command name

CMD_QUERY_POWER

Description

The command reads the power state.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

0

0

end flag

Byte String (HEX)

A5

5A

59

00

00

00

00

00

00

00

B9

9B

Results

see MMCA_QUERY_POWER function in DLL description.

Command name

CMD_QUERY_STATE

Description

The command reads the state.

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

0

0

end flag

Byte String (HEX)

A5

5A

5A

00

00

00

00

00

00

00

B9

9B

Results

see MMCA_QUERY_STATE function in DLL description.

Command name

CMD_QUERY_SPECTRA

Description

The command reads the spectrum data.
n: No. of 1. channel and buffer control (Bit 15 ... 12)
c: Compress factor (1 ... 32)

Command syntax

Format

integer

integer

integer

long

integer

Parameter

preamble

command

n

C

end flag

Byte String (HEX)

A5

5A

5B

00

n

l

n

h

c

l

c

h

00

00

B9

9B

Used setup value

channels, lld, uld

Results

see MMCA_QUERY_SPECTRA function in DLL description.

Command name

CMD_QUERY_IMAGE

Description

The command read the MMCA spectrum image (120x256).
n: No. of 1. Channel
c: Compress factor (1 ... 32)
v: Vertical full scale (see MCA_VFS in mca_comm.h)

Command syntax

Format

integer

integer

integer

integer

integer

integer

Parameter

preamble

command

n

C

v

end flag

Byte String (HEX)

A5

5A

5C

00

n

l

n

h

c

l

c

h

v

h

v

h

B9

9B

Used setup value

channels, lld, uld

Results

see MMCA_QUERY_IMAGE function in DLL description.