13miscellaneous, 13 miscellaneous – GBS Elektronik MCA166-USB User Manual
Page 14
![background image](/manuals/55061/14/background.png)
MCA166-USB
3.13 Miscellaneous
For communication and evaluation some libraries for MS Windows and Linux are
available. This may help the make own application programs or to integrate the
MCA166-USB in an existing system.
4 MCA166-USB parameters for use with different detectors
This table shall just give an overview of possible settings useful with the MCA166-USB. It
is far from being complete. For actual settings with your detector, see the detectors
manual.
Table 2: Parameters for different detectors
Type
Model
Manufa
ct.
HV
(V)
+
12
V
-1
2V
+
24
V
-
24
V
Input s
ignal
pola
rity
Chann
els
Ga
in for
2
M
eV
full r
ange
Shap
ing
tim
e (µs)
CdZnTe SDP/Z/60
Ritec
+300
x
x -
-
neg
1024 90
1
CdZnTe SDP/Z/20
Ritec
+500
x
x -
-
neg
1024 80
1
CdZnTe CZT500
Ritec
+600
x
x -
-
pos
1024 128
1
NaI
Scintiblock
Crismatec +500
pos
512
7.2
1
NaI
25825
Scionix
+550
x
x -
-
pos
512
90
1
NaI
40*40
Amcrys-h -800
x
x
pos
512
14
1
CsI
Sc1010
Eurorad
-
x
x -
-
neg
256
160
1
HPGe
GMX-20190-S Ortec
-3000
x
x x
x
neg
4096 18
2
HPGe
GEM 15-190
Ortec
+3000 x
x x
x
pos
4096 24
2
HPGe
GR 2018
Canberra -3000
neg
4096 11
2
HPGe
GL 0310
Canberra -2000
x
x x
-
neg
4096 2.3
2
Si
planar
7865S
Ortec
-1000
neg
4096 1.9
2
HPGe
EGC 30-190R Eurisys
-3000
pos
4096 52
2
HPGe
EGPC 25-185 Eurisys
+2000
neg
4096 35
2
14