GBS Elektronik MCA166-USB User Manual
Page 27
Setup information - continued
$MCS_CHANNELS:
Multi Channel Scaler (MCS) Setup
4096
Channels
$MCS_INPUT:
MCS input
Extern TTL
Extern TTL (external TTL signal),
Input Rate (count rate from the internal amplifier / fast
discriminator)
$MCS_TIME:
MCS time per channel
10
Time in msec , 0.01 – 655.35 sec
$MCS_SWEEPS:
Number of sweeps
0
0 .... 65535
0
MCS Repeat Mode Type
$MODE:
Operation Mode
MCA
MCA or MCS
$MCA_REPEAT:
Number of sweeps
1
0 .... 65535
0
MCA Repeat Mode Type
$TDF
Dead Time Correction Factor
800
100 ... 3000, default 800 nsec
$POWER_STATE:
Power state at the end of the measurement
I+12= 8mA
Input current of the DC-DC converters for +12V preamplifier
power supply from the battery
I-12= 3mA
Input current of the DC-DC converters for -12V preamplifier
power supply from the battery
I+24= 0mA
Input current of the DC-DC converters for +24V preamplifier
power supply from the battery
I-24= 1mA
Input current of the DC-DC converters for -24V preamplifier
power supply from the battery
IBAT= 135mA
Total current drawn from battery
IHV = 14mA
Input current of the HV DC-DC converter drawn from the battery
ICHR= 0mA
External charger current
UBAT=7900mV
Battery Voltage
UHVs= 418mV
Control voltage of the HV module
$COUNTS:
Integral counts
8370252
Sum of all input counts (from the internal amplifier / fast
discriminator) of the whole measurement
$PD_COUNTS:
Integral peak detector counts
5113594
Sum of all input counts from the peak detector
$RT:
Real time [s]
203
$DT:
Dead time [ms]
883027
$BT:
Busy time [ms] of the ADC
64042
$STAB_OFFSET:
Current offset
315
$STAB_OFFSET_MIN:
Minimal offset
310
$STAB_OFFSET_MAX:
Maximal offset
408
$STAB_COUNTER:
Stabilization cycles
19
27