GBS Elektronik MCA166-USB User Manual
Page 67
MCA Detector Power Commands:
Command name
CMD_SET_PREAMPLIFIER_POWER
Description
The command sets the preamplifier power(pp).
0X80
-24V
on
0X40
+24V
on
0X20
-12V
on
0X10
+12V
on
0XF0
all on
0X00
all off
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
pp
0
end flag
Byte String (HEX)
A5
5A
4E
00
pp
00
00
00
00
00
B9
9B
Used setup value
None
Affected setup value
preamplifier_power
MCA state
STATE_READY
Command name
CMD_SET_BIAS
Description
The command sets the detector high voltage
and controls the HV-inhibit-input.
hv:
0 ... 3000
i = 0
Inhibit off
i = 1
”Canberra- or DFG-mode”, HV shut down if inhibit input < 0.5V
i =-1
”Ortec-mode”, HV shut down if inhibit input = +5V
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
hv
I
end flag
Byte String (HEX)
A5
5A
4F
00
hv
l
hv
h
i
l
i
h
00
00
B9
9B
Used setup value
None
Affected setup value
detector_bias, stab_state
MCA state
STATE_READY
42 The high voltages will ramp up or down by the MCA166-USB.
67