beautypg.com

GBS Elektronik MCA166-USB User Manual

Page 64

background image

MCA166-USB

Command name

CMD_SET_PUR

Description

The command sets the pile up rejection.

Command syntax

pur = 0 turns PUR off
pur

0 turns PUR on

Format

integer

integer

integer

long

integer

Parameter

preamble

command

pur

0

end flag

Byte String (HEX)

A5

5A

53

00

pur

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

mca_pur

MCA state

STATE_READY

Command name

CMD_SET_FAST

Description

This command set the fast discriminator threshold (0 ... 2499, default 400).

Command syntax

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

fast

0

end flag

Byte String (HEX)

A5

5A

50

00

fast

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

Fast threshold

MCA state

STATE_READY

Command name

CMD_SET_SLOW

Description

This command set the slow discriminator threshold (0 ... 2499, default 400).

Command syntax

Format

integer

integer

integer

long

integer

Parameter

Preamble

command

slow

0

end flag

Byte String (HEX)

A5

5A

51

00

slow

00

00

00

00

00

B9

9B

Used setup value

None

Affected setup value

Slow threshold

MCA state

STATE_READY

64