Command name
CMD_SET_PZC
Description
The command performs the Pole Zero Cancellation.
pv: 0 ... 2499
PZC control voltage
pm = 0
set PCZ only
pm > 0
set PZC and measure PZC offset
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
pm
Pv
end flag
Byte String (HEX)
A5
5A
58
00
pm
l
pm
h
pv
l
pv
h
00
00
B9
9B
Used setup value
None
Affected setup value
pzc_value
Results
If pm > 0 the function returns with the number of counts and the - PZC offset.
MCA state
STATE_READY
Command name
CMD_SET_PZC_TIME_OFFSET
Description
The command sets the time offset for Pole Zero Cancellation.
t1: 0 ... 31
Time offset for low shaping time
t2: 0 ... 31
Time offset for high shaping time
Command syntax
Format
integer
integer
integer
long
integer
Parameter
preamble
command
t1
t2
end flag
Byte String (HEX)
A5
5A
60
00
t1
00
t2
00
00
00
B9
9B
Used setup value
None
Affected setup value
pzc_dtc_1_offset, pzc_dtc_3_offset
MCA state
STATE_READY