beautypg.com

GBS Elektronik MCA166-USB User Manual

Page 39

background image

MCA Mode Commands:

Name

MMCA_SET_MODE_MCA

Description

The command sets the acquire mode to Multi-Channel Analyser

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_MODE_MCS

Description

The command sets the acquire mode to Multi-Channel Scaler

Parameter

-

Results

-

Return Code

see MMCA_RESET command

MCA Clear Commands:

Name

MMCA_CLEAR_MEMORY

Description

The acquisition is stopped and the spectrum is cleared

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_CLEAR_TIME

Description

The acquisition is stopped and the dead time together with the real time are set to 0

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_CLEAR_ROI

Description

The acquisition is stopped and the ROI limits are set to LLD and ULD

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_CLEAR_ALL

Description

The command is a combination of MMCA_CLEAR_ROI, MMCA_CLEAR_MEMORY and
MMCA_CLEAR_TIME

Parameter

-

Results

-

Return Code

see MMCA_RESET command

39