GBS Elektronik MCA166-USB User Manual
Page 40
MCA166-USB
MCA Measurement Setup Commands:
Name
MMCA_SET_PRESET_NONE
Description
None preset
Parameter
-
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_PRESET_LIVE_TIME
Description
The command sets the time for measurement (dead time corrected).
1. Parameter
Live time
integer value
1 ... 65535 sec
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_PRESET_REAL_TIME
Description
The command sets the time for measurement
1. Parameter
Real time
integer value
1 ... 65535
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_PRESET_INTEGRAL
Description
The command sets the value for the ROI integral
1. Parameter
Integral
long value
1 ... 4294967295
Results
-
Return Code
see MMCA_RESET command
Name
MMCA_SET_PRESET_AREA
Description
The command sets the value for the ROI area
1. Parameter
Area
long value
1 ... 4294967295
Results
-
Return Code
see MMCA_RESET command
40