MCA166-USB
Name
MMCA_QUERY_UF6_INFO
Description
The command reads current dead time, real time the integral, begin and end of the UF6 ROIs
Parameter
-
Result array
Real time
long value
Byte Offset 4
Dead time [msec]
long value
Byte Offset 8
Integral ROI 1
long value
Byte Offset 12
Integral ROI 2
long value
Byte Offset 16
Integral ROI 3
long value
Byte Offset 20
Begin ROI 1
long value
...
...
long value
Byte Offset 40
End ROI 3
long value
Return Code
see MMCA_RESET command
Name
MMCA_QUERY_UF6_ROIS
Description
The command reads the begin and end of the 3 ROIs
Parameter
-
Result array
Begin ROI 1
long value
Byte Offset 4
End ROI 1
long value
Byte Offset 8
Begin ROI 2
long value
Byte Offset 12
End ROI 2
long value
Byte Offset 16
Begin ROI 3
long value
Byte Offset 20
End ROI 3
long value
Return Code
see MMCA_RESET command
Name
MMCA_QUERY_CENTROID
Description
The command reads peak centroid of the specified ROI
1. Parameter
Peak ROI begin
integer value
LLD<= begin < end
2. Parameter
Peak ROI end
integer value
begin < end <= ULD, (end-begin) < 250
Result
Peak centroid
float value
Return Code
see MMCA_RESET command
Name
MMCA_QUERY_ON_LINE
Description
The command checks the communication between the MCA and the PC.
Parameter
-
Results
-
Return Code
see MMCA_RESET command