MCA Hardware Setup Commands:

Name

MMCA_SET_GAIN

Description

The command sets the amplifier coarse and fine gain.

1. Parameter

Coarse gain

integer value

2, 5, 10, 20, 50, 100, 200, 500 or 1000

2. Parameter

Fine gain

15

integer value

5000 ... 15000 (10000 if Coarse Gain=1000)

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_MCA_INPUT_AMPLIFIER_POS

Description

The command sets the amplifier input polarity to positive.

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_MCA_INPUT_AMPLIFIER_NEG

Description

The command sets the amplifier input polarity to negative.

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_MCA_INPUT_DIRECT_POS

Description

The command sets the ADC input to direct (0 to +3V input range) and PUR to off.

Parameter

-

Results

-

Return Code

see MMCA_RESET command

Name

MMCA_SET_MCA_INPUT_DIRECT_NEG

Description

The command sets the ADC input to direct (0 to -3V input range) and PUR to off.

Parameter

-

Results

-

Return Code

see MMCA_RESET command

15 It is the real fine gain value multiplied with 10000.