4 configuration of the indicator, Configuration of the indicator – WIKA DI35 User Manual
Page 27

6 Commissoning, operation
WIKA operating instructions digital indicator model DI35-M
27
6.4 Configuration of the indicator
Functional diagram of programming via key pad:
Description of the program numbers
In the display, the program numbers (PN) are shown, right-justified, as a 3-
digit number with a P in front of them:
Programming procedure
The entire programming of the DI35 is done by the steps described below.
Change to programming mode
Push the [P] key to change into programming mode. The unit goes to the
lowest available program number.
If the programming lock is activated, the key must be pushed for at least 1
second.
Example:
Displaying of program number 0, after pushing
the [P] key.
Change between program numbers
To change between individual program numbers, hold the [P] key down and
push the
[▲] key for changing to a higher program number or the [▼] key for
changing to a lower number. By keeping the keys pushed, e.g. [P] &
[▲], the
display will begin, after approx. 1 second, to automatically run through the
program numbers.