3 function and operation description, 1 operation, 2 switching on – WIKA DI35 User Manual
Page 20: Function and operation description

6 Commissoning, operation
20
WIKA operating instructions digital indicator DI35-M
6.3 Function and operation description
6.3.1 Operation
The DI35-M has 4 keys ([P], [
], [
], [O]), with which you can parameterise
and call up various functions during operation.
Name
Description
Program key [P]
With the program key, you can call up the
programming mode or perform various functions in the
programming mode.
DOWN key [
▼]
With the DOWN (decrease) key, you can call up the
MIN memory or alter parameters in the programming
mode.
UP key [
▲]
With the UP (increase) key, you can call up the MAX
memory or alter parameters in the programming mode.
Multi-function [O]
Multi-function for activating TARA function and HOLD
function; reset for MIN/MAX permanent
6.3.2 Switching on
Before switching on, check all the electrical connections to make sure they
are correct. On completion of the installation, the device can be switched on
by applying the supply voltage.
During the switching-on process, a segment test is performed for approx. 1
second, whereby all LED on the front (including setpoint LED) are triggered.
After this, the type of software is indicated for approx. 1 second and then,
also for 1 second, the software version. After the starting procedure, the unit
changes to operation/display mode.