Cpt6200 pressure transmitter specific details – WIKA CPH6200 User Manual
Page 24
Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
GB
11221780.1.3 01/2009 GB/D/F/E
WIKA Operating Instructions Hand-Held Pressure Indicator
24
CPT6200 Pressure Transmitter specific details:
Pressure connection
G½ B; {flush diaphragm (G1 for 0.1 up to 1.6 bar) or various connection
adapters on request}
Material wetted parts
Wetted parts stainless steel, case stainless steel;
flush diaphragm version: stainless steel {Hastelloy C4};
O-Ring
2)
: NBR {FKM/FPM or EPDM}
Internal transmission fluid
Synthetic oil, (only for pressure ranges up to 16 bar or flush
diaphragm), {Halocarbon oil for oxygen applications}
3)
;
{Listed by FDA for food industry}
One year stability
0.2 % of span at reference conditions
Permissible
- medium temperature
3)
-25 ... +100 °C
- ambient temperature
0 ... 50 °C
- storage temperature
-40 ... +80 °C
Compensated range
0 ... 70 °C
Temperature coefficients
- mean TC of zero signal
0.2 %/10 K
- mean TC of span
0.2 %/10 K
Connection to CPH6200
via 1 m cable (plug and play); on request: up to 5 m
Weight
approx. 220 g
2) O-ring made of Viton or EPDM for flush diaphragm with integrated cooling element.
3) The oxygen version must not be used with medium temperatures higher than 60 °C.
Cannot be manufactured for absolute pressure ranges < 1 bar abs.
{} Items in curved brackets are optional extras at an additional price.