Calibration services, Specifications, Cph6200 digital instrument specific details – WIKA CPH6200 User Manual
Page 23
Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
GB
11221780.1.3 01/2009 GB/D/F/E
WIKA Operating Instructions Hand-Held Pressure Indicator
23
Input
1: for CPH6200-S1;
2: for CPH6200-S2
Pressure range in bar
0.1
0.16
0.25
0.4
0.6
1.0
1.6
2.5
4.0
6.0
Overpressure safety in bar 1
1.5
2
2
4
5
10
10
17
35
Burst pressure in bar
2
2
2.4
2.4
4.8
6
12
12
20.5
42
Resolution in mbar
0.1
1
Pressure range in bar
10
16
25
40
60
100
160
250
400
600
1000
Overpressure safety in bar 35
80
50
80
120
200
320
500
800
1200
1500
Burst pressure in bar
42
96
250
400
550
800
1000
1200
1700
2400
3000
Resolution in bar
0.01
0.1
1
Type of pressure
gauge {absolute 0.25 up to 16 bar & vacuum on request}
Accuracy of the measu-
ring chain
0.2 % FS +/- 1 digit at reference temperature of 20 °C (optional 0.1 %)
Calibration*
factory calibration certificate (optional: DKD calibration certificate)
Display
large LCD display for display of 2 values and additional information
Range of display
max. -19999 up to 99999, depending on sensors used
Units
can be selected from the following: mbar, bar, Pa, kPa, MPa, mmHg, inHg, psi
(dependent on range)
CPH6200 Digital Instrument specific details:
Functions via key press
Min-, Max-memory, Hold, Tare, Offset-correction, Logger (Start/Stop)
Functions via Menu
Min-, Max-alarm (acoustic**/visual), Sea-level (barom. air pressure), Power-Off-
Function, measuring rate: 4/sec (“slow“); > 1000/sec (“fast“); > 1000/sec unfil-
tered (“peak-detect“) [via “peak-detect“ and min-/max-memory: pressure peaks
of 1.5 msec can be detected]
Data logger
- discrete value-logger: up to 99 data points (incl. time-stamp) via key press
- cyclic-logger: automatic logging of up to 10000 values incl. time.
cycle-time: selectable between 1 ... 3600 seconds
Interface
1)
(serial)
RS-232 or USB interface via special interface cables
Power supply
9V-zinc-carbon battery, alternative**: 9V rechargeable battery or mains supply
Power consumption
Cycle-time slow: < 1.6 mA, fast: < 7.0 mA, Low-Power-Logger-Fct.: < 0.3 mA
Working temperature
0 ... 50 °C
Air humidity
0 ... 95 % relative humidity without moisture condensation
Storage temperature
-20 ... +70 °C
Housing
Impact-resistant ABS, membrane keyboard, transparent panels (CPH62I0 with
cover)
Weight
Approx. 160 g
EMV- / CE-conformity
The CPH6200 device corresponds to the essential protection requirements estab-
lished in the regulations of the council for the approximation of legislation of the
member countries regarding electromagnetic compatibility (89/336/EWG).
1) For the intrinsically safe version, the use of the interface is not permitted within an explosive atmosphere.
*) Calibrated in vertical mounting position with lower pressure connection.
**) Not available for the intrinsically safe version CPH62I0.
{} Items in curved brackets are optional extras at an additional price.
7. Calibration services
DKD-certificates - other certificates:
If the instrument requires calibration, we recommend returning it, along with its associ-
ated sensors, to the manufacturer. Only the manufacturer is able to verify the instrument‘s
basic settings and configuration, and if necessary correct them.
8. Specifications