Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2

GB

11221780.1.3 01/2009 GB/D/F/E

WIKA Operating Instructions Hand-Held Pressure Indicator

23

Input

1: for CPH6200-S1;

2: for CPH6200-S2

Pressure range in bar

0.1

0.16

0.25

0.4

0.6

1.0

1.6

2.5

4.0

6.0

Overpressure safety in bar 1

1.5

2

2

4

5

10

10

17

35

Burst pressure in bar

2

2

2.4

2.4

4.8

6

12

12

20.5

42

Resolution in mbar

0.1

1

Pressure range in bar

10

16

25

40

60

100

160

250

400

600

1000

Overpressure safety in bar 35

80

50

80

120

200

320

500

800

1200

1500

Burst pressure in bar

42

96

250

400

550

800

1000

1200

1700

2400

3000

Resolution in bar

0.01

0.1

1

Type of pressure

gauge {absolute 0.25 up to 16 bar & vacuum on request}

Accuracy of the measu-
ring chain

0.2 % FS +/- 1 digit at reference temperature of 20 °C (optional 0.1 %)

Calibration*

factory calibration certificate (optional: DKD calibration certificate)

Display

large LCD display for display of 2 values and additional information

Range of display

max. -19999 up to 99999, depending on sensors used

Units

can be selected from the following: mbar, bar, Pa, kPa, MPa, mmHg, inHg, psi
(dependent on range)

CPH6200 Digital Instrument specific details:

Functions via key press

Min-, Max-memory, Hold, Tare, Offset-correction, Logger (Start/Stop)

Functions via Menu

Min-, Max-alarm (acoustic**/visual), Sea-level (barom. air pressure), Power-Off-
Function, measuring rate: 4/sec (“slow“); > 1000/sec (“fast“); > 1000/sec unfil-
tered (“peak-detect“) [via “peak-detect“ and min-/max-memory: pressure peaks
of 1.5 msec can be detected]

Data logger

- discrete value-logger: up to 99 data points (incl. time-stamp) via key press
- cyclic-logger: automatic logging of up to 10000 values incl. time.

cycle-time: selectable between 1 ... 3600 seconds

Interface

1)

(serial)

RS-232 or USB interface via special interface cables

Power supply

9V-zinc-carbon battery, alternative**: 9V rechargeable battery or mains supply

Power consumption

Cycle-time slow: < 1.6 mA, fast: < 7.0 mA, Low-Power-Logger-Fct.: < 0.3 mA

Working temperature

0 ... 50 °C

Air humidity

0 ... 95 % relative humidity without moisture condensation

Storage temperature

-20 ... +70 °C

Housing

Impact-resistant ABS, membrane keyboard, transparent panels (CPH62I0 with
cover)

Weight

Approx. 160 g

EMV- / CE-conformity

The CPH6200 device corresponds to the essential protection requirements estab-
lished in the regulations of the council for the approximation of legislation of the
member countries regarding electromagnetic compatibility (89/336/EWG).

1) For the intrinsically safe version, the use of the interface is not permitted within an explosive atmosphere.
*) Calibrated in vertical mounting position with lower pressure connection.
**) Not available for the intrinsically safe version CPH62I0.
{} Items in curved brackets are optional extras at an additional price.

7. Calibration services

DKD-certificates - other certificates:
If the instrument requires calibration, we recommend returning it, along with its associ-
ated sensors, to the manufacturer. Only the manufacturer is able to verify the instrument‘s
basic settings and configuration, and if necessary correct them.

8. Specifications