Fault and system messages – WIKA CPH6200 User Manual
Page 22
Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2
GB
11221780.1.3 01/2009 GB/D/F/E
WIKA Operating Instructions Hand-Held Pressure Indicator
22
Display
Meaning
Remedy
Low battery power, device will only work
for a short while longer
Replace battery
Battery flat
Replace battery
Mains operation without battery: wrong
voltage
Check power supply, change it if neces-
sary
SEnS
Erro
or Err.9
No sensor connected
Switch off device and connect sensor
Connected sensor or device faulty
If second sensor available, check if
device is ok. Return defective device/
sensor to manufacturer for repair
Value extremely out of measuring range
Check: pressure not within sensor range?
- - - -
- - - -
Logger data is being read by the serial
interface
When the data transfer is completed the
device will automatically return to normal
measuring display, no remedy necessary
No display
or strange
symbols,
device is not
responding to
input
Battery flat
Replace battery
Mains operation without battery: wrong
voltage or polarity
Check power supply, change it if neces-
sary
System error
Disconnect battery or power supply, wait
a while, re-connect
Instrument faulty
Return to manufacturer for repair
Err.1
Measured value above allowable range
Check: Is the pressure outside the
sensor‘s range? -> measured value to
high!
Sensor faulty
Return to manufacturer for repair
Err.2
Measured value below allowable range
Check: Is the pressure outside the
sensor‘s range? -> measured value to
low!
Sensor defective
Return to manufacturer for repair
Err.3
Display range overflow
Check: value above 9999 -> to high to be
displayed!
Err.4
Display range underflow
Check: value below ‚-2000‘ (Tare?) ->
to low!
Err.7
System error
Return to manufacturer for repair
Err.11
Value coul not be calculated
Choose different unit
Calculation overflow happened
Choose different unit
6. Fault and system messages