Fault and system messages – WIKA CPH6200 User Manual

Page 22

background image

Hand-Held Pressure Indicator
CPH6200-S1 / CPH6200-S2

GB

11221780.1.3 01/2009 GB/D/F/E

WIKA Operating Instructions Hand-Held Pressure Indicator

22

Display

Meaning

Remedy

Low battery power, device will only work
for a short while longer

Replace battery

Battery flat

Replace battery

Mains operation without battery: wrong
voltage

Check power supply, change it if neces-
sary

SEnS
Erro

or Err.9

No sensor connected

Switch off device and connect sensor

Connected sensor or device faulty

If second sensor available, check if
device is ok. Return defective device/
sensor to manufacturer for repair

Value extremely out of measuring range

Check: pressure not within sensor range?

- - - -
- - - -

Logger data is being read by the serial
interface

When the data transfer is completed the
device will automatically return to normal
measuring display, no remedy necessary

No display
or strange
symbols,
device is not
responding to
input

Battery flat

Replace battery

Mains operation without battery: wrong
voltage or polarity

Check power supply, change it if neces-
sary

System error

Disconnect battery or power supply, wait
a while, re-connect

Instrument faulty

Return to manufacturer for repair

Err.1

Measured value above allowable range

Check: Is the pressure outside the
sensor‘s range? -> measured value to
high!

Sensor faulty

Return to manufacturer for repair

Err.2

Measured value below allowable range

Check: Is the pressure outside the
sensor‘s range? -> measured value to
low!

Sensor defective

Return to manufacturer for repair

Err.3

Display range overflow

Check: value above 9999 -> to high to be
displayed!

Err.4

Display range underflow

Check: value below ‚-2000‘ (Tare?) ->
to low!

Err.7

System error

Return to manufacturer for repair

Err.11

Value coul not be calculated

Choose different unit

Calculation overflow happened

Choose different unit

6. Fault and system messages