a. TPC “Calibration” Module Procedure
This procedure assumes the front panel is removed.
1. Ensure power is OFF.
2. Remove the sensor leads from the Temperature
Module.
3. Remove the 0-5VDC leads from the Temperature
Module.
4. Press TP Calibration Module (TPC) firmly into sensor
input connectors as shown in figure III.C.5.
5. Attach the DVM to the 0-5VDC output on the TPC
module. See figure III.C.5.
6. Turn power ON. The DVM should indicate 0.0 volts.
7. If not, adjust the calibration control marked “ZERO”
(see figure III.C.1 for location) until the DVM indicates
0.0 volts as shown in figure III.C.5.
8. Press switch on the TPC. See figure III.C.6.
9. The DVM should indicate 5.0 volts. See figure III.C.6.
10. If not, while pressing the TPC Module switch, adjust
the calibration control marked “FS” (see figure III.C.1 for
location) until the DVM indicates 5.0 volts as shown in
figure III.C.6.
11. Press the front panel TEMPERATURE select switch
and the TPC Module switch, the display should
indicate 199.9 as shown in figure III.C.7.
12. If not adjust the calibration control marked “DIS” (see
figure III.C.1 for location) until the reading is 199.9 as
shown in figure III.C.7.
13. Calibration is complete.
14. Turn power OFF.
15. Remove TPC Module by pressing on each Phoenix
connector release lever*.
16. Reconnect sensor leads as labeled.
17. Reconnect 0-5VDC output leads as labeled.
18. Continue or reinstall the front panel and tightly secure
four (4) retaining screws, see REASSEMBLY below.
*CAUTION: The sensor input and 0-10VDC output connectors
require only a small screwdriver or a pen to push on the release
levers. The release levers may be broken or damaged if not
pushed straight toward the CB. DO NOT push the release levers
sideways.
b. Precision Resistor Calibration Procedure
This procedure assumes the front panel is removed.
1. Ensure power is OFF.
2. Remove the sensor leads from the Temperature Module.
3. Install ZERO resistor (1000.0Ω) across sensor terminals
BL and OR. See figure III.C.8.
4. Install JUMPER between BR & BL as shown in
figures III.C.8 & III.C.9.
5. Attach the DVM to the 0-5VDC output. See
figures III.C.8 & III.C.9.
6. Turn power ON. The DVM should indicate 0.0 volts.
7. If not, adjust the calibration control marked “ZERO”
(see figure III.C.1 for location).
8. Install SPAN resistor (1758.56Ω) across sensor
terminals BL and OR. See figure III.C.9.
9. The DVM should indicate 5.0 volts.
10. If not, adjust the calibration control marked “FS” (see
figure III.C.1 for location) until the DVM indicates 5.0
volts as shown in figure III.C.9.
11. Press the front panel TEMPERATURE select switch,
the display should indicate 199.9 as shown in
figure III.C.10.
12. If not adjust the calibration control marked “DIS” (see
figure III.C.1 for location) until the reading is 199.9.
13. Calibration is complete.
14. Turn power OFF.
15. Remove DVM from 0-5VDC.
16. Remove resistor and jumper by pressing on each
Phoenix connector release lever*.
17. Reconnect sensor leads as labeled.
18. Reconnect 0-5VDC output leads as labeled.
Figure III.C.5
DV
-
+
0.0
Press TPC Module into
TPO connector as shown.
TPO connector as shown.
Connect DVM leads as shown.
BR BL OR - +
TPC Module
PRESS BUTTON FOR
199.9ºC/5.0VDC
+
OUTPUT -
ORANGE
BLUE
BROWN
0-5VDC +
OUTPUT -
ORANGE
BLUE
BROWN
MYRON L
COMPANY
Figure III.C.6
Press TPC Module into
TPO connector as shown.
Connect DVM leads as shown.
DV
-
+
5.0
BR BL OR - +
MYRON L
COMPANY
0-5VDC +
OUTPUT -
ORANGE
BLUE
BROWN
0-5VDC +
OUTPUT -
ORANGE
BLUE
BROWN
TPC Module
PRESS BUTTON FOR
199.9ºC/5.0VDC
Figure III.C.7
199.9
FRONT PANEL DISPLAY