beautypg.com

Dow FROTH-PAK Refill Systems User Manual

Page 27

background image

dow Building Solutions

9 .4 System Pressurization

1 . Open the nitrogen bottle valve by turning the knob

on top of the cylinder counter-clockwise . If a hissing

sound occurs, further tighten the nitrogen regulator

valve stem with an adjustable wrench .

2 . Verify that nitrogen cylinder pressure using the center

gauge of the nitrogen regulator assembly . The system

WIll nOT operate efficiently if the pressure falls below

500 psi . If below 500 psi, replace the nitrogen cylinder .

3 . Turn the left regulator stem clockwise to the proper

pressure setting of the “a” tank (120-160psi – Do not

go over 225psi) . If a hissing sound occurs, further

tighten the yellow-coded nitrogen line connection to

the nitrogen regulator assembly and/or the nitrogen

intake connection on the “a” tank using an adjustable

wrench . If the left gauge fails to indicate a pressure

setting, contact your Dow Sales Representative .

4 . Turn the right regulator stem clockwise to the proper

pressure setting of the “B” tank (120-160psi – Do not go

over 225psi) . If a hissing sound occurs, further tighten the

yellow-coded nitrogen line connection to the nitrogen

regulator assembly and/or the nitrogen intake valve

connection on the “B” tank using an adjustable wrench .

If the right gauge fails to indicate a pressure setting,

contact your Dow Sales Representative .

5 . Slowly open the nitrogen intake valve of each tank .

The valve handles should be parallel with the nitrogen lines .

6 . Slowly turn on the chemical valve of each tank and the on-off

valve of each chemical line . The valve handles should be

parallel to the chemical lines . Verify that the filter connection

of each tank is tight and that no chemical leaks are present .

9 .5 System Purging and Testing

1 . Dispense chemicals in an appropriate container .

This is to verify proper chemical flow .

2 . Clean any chemical from the InSTa-FlO™ Gun face

using a rag .

3 . Insert an unused nozzle with the key slot down . Push in

firmly until the nozzle ejector is seated over the back rim of

the nozzle .

4 . Dispense foam to verify proper chemical mixing . If improper

mixing or poor foam quality, refer to the Troubleshooting

section, Section 10 of this manual .

Section 10

Troubleshooting

The FROTH-PaK™ Refill System is virtually maintenance-free and

many problems can be corrected through simple troubleshooting
techniques . When troubleshooting, confirm that the system is pressur-
ized correctly, and that all nitrogen and chemical valves are in the open
position . One closed valve can cause a system shut-down .

10 .1 Temperature

Chemical temperature can affect foam quality . If the chemical tem-
perature is below 65

°

F (18

°

C), poor quality foam may result . The ideal

temperature range is 70

°

-90

°

F (21

°

-32

°

C) .

10 .2 Foam Color and Quality

Changes in foam quality indicate an off-ratio foam . If the foam is crusty
after curing, the foam is “a” rich and a blockage of “B” chemical exists .

If the foam is mushy, and remains soft after two minutes, the foam is “B”
rich and a blockage of “a” chemical exists .

To identify and correct the problem, initiate the following procedure:

1 .

dO nOT breathe vapor or spray . Proper PPE and ventilation

are required . See product SDS and Section 2 of this manual

for further information . Follow all precautions for product .

2 . Remove the used nozzle by pushing down the nozzle

ejector . use only in well ventilated areas . With insuf-

ficient ventilation, wear suitable respiratory protection .

19