beautypg.com

Vigilant – Maximum Vigilant User Manual

Page 2

background image

VIGILANT

OPERATION

TROUBLE SHOOTING

ADDITIONAL INFORMATION

To reset the gust register needle,
turn the knob counter-clockwise.

The Vigilant is a self-powered wind speed measuring instrument. It does not need batteries
or an external power supply. In normal operation, it requires no maintenance.

Turning the knob counter-clockwise will eventually contact the gust register.
The Gust Register should not be set under 15 MPH - this gives the needle
room to operate at low speeds. If the Gust Register is set below 15 MPH
turn the knob clockwise until the needle registers greater than 15 MPH.

After setting the Gust Register turn the knob clockwise
until the reset arm is parked near the six o'clock position.
Failure to do so may damage your meter.

1.

Disconnect the wires from the back of the brass read-out.

2.

Connect an ohm meter to the wires.

3.

The ohm meter should indicate between 1,000 and 2,000 ohms.

4.

If there is a good reading on the ohm meter, then the brass read-out

is faulty. If the reading is bad, then the sensor or wire is faulty.

1.

Disconnect the wires from the back of the brass read-out.

2.

Attach a low range AC Voltmeter to the wires. If the sensor is operating

properly, you will achieve these approximate readings (using 20,000 ohms
per volt): 8-9 MPH = 0.28 Vac rms, 17 MPH = 0.56 Vac rms,
51 MPH = 1.78 Vac rms, 102 MPH 3.67 Vac rms

3.

If the sensor delivers these approximate readings, then the brass read-out

is faulty. If the sensor does not produce these readings, then either the
sensor or the wire is faulty.

WIRE

GAUGE

MAXIMUM FEET

WITHOUT RECALIBRATION

WIRE

GAUGE

MAXIMUM FEET

WITHOUT RECALIBRATION

24

22

20

19

18

300'

500'

750'

1000'

1250'

(supplied with instrument)

Cables can be shortened or lengthened without affecting accuracy

If there is

no wind

If there is

wind

16

14

12

10

2000'

3200'

5000'

8000'

Samuel Barnett Boulevard

New Bedford, MA 02745

(508) 995-2200

Maximum Instruments are accurate and reliable. Most problems that occure are due
to loose or corroded connections. If, after checking the connections, there is still a
problem, determine if the problem is with the sensor or the brass read-out.