beautypg.com

3 datalogger programming, 1 sensor sensitivity, 2 example programs – Campbell Scientific CNR4 Net Radiometer User Manual

Page 29: Datalogger programming, Sensor sensitivity, Example programs, Example 1, cr1000 program using differential, Measurements, 3 datalogger pro ming, Gram

background image

CNR4 Net Radiometer

7.3 Datalogger Pro

ming

single-ended measurement. If differential channels are not available, single-

ptability of a single-ended

ed by simply comparing the results of single-

y,

7.3.1 Sensor Sens

rate

nufacturer. They show the sensor serial number and

sensitivity values for four individual probes: one copy for pyranometers, and

y for pyrgeometers. The serial number and sensitivity values are

on a label affixed to the bottom of the sensor. If you choose to

CNF4

.

7.3.2 Example Pro

7.3.2.1 Example 1, CR

ur differential channels to measure the four radiation

outputs, one excitation channel, and one single-ended channel to measure the

he program measures the sensors every 1 second, performs the

ssing of the data, and stores the following processed data to a data

es the raw time-series

ometer up)

Average long-wave radiation (pyrgeometer down)

temperature (degrees C)

(Kelvin)

eter up)

wn)

gram

The CNR4 outputs four voltages that typically range from 0 to 15 mV for the
pyranometers, and

± 5 mV for the pyrgeometers. A differential voltage

measurement is recommended because it has better noise rejection than a

ended measurements can be used. The acce
measurement can be determin
ended and differential measurements made under the same conditions.

Additionally, one voltage excitation channel and one single-ended analog
channel are required to make the temperature measurement of the sensor bod
using the thermistor.

itivity

The CNR4 comes with four different sensor sensitivity values for four sepa
probes. The CNR4 sensor comes with two copies of its ‘Certificate of
Calibration’ by the ma

another cop
also shown
attach the CNF4 heater/ventilator unit to the CNR4, the label showing the
serial number and sensitivity values will be covered. After attaching the
heater/ventilator, affix the extra label to the bottom of the CNF4 in a visible
location. The extra label containing the serial number and sensitivity values is
supplied with the purchase of the CNR4. Please refer to Appendix B, CNF4
Heater/Ventilator
, for more details.

The sensor sensitivity is in

μV/(W/m

2

). This needs to be converted into

(W/m

2

)/mV to be used as a multiplier parameter inside the datalogger program

To convert the units, divide the sensor sensitivity value into 1000. For
example, if the sensitivity is 7.30

μV/(W/m

2

), the multiplier is 1000/7.3 =

136.99 (W/m

2

)/mV.

grams

1000 Program Using Differential Measurements

Example 1 requires fo

thermistor. T
online proce
table called cnr4_data once every 60 minutes. It also stor
data from CNR4 to data table called cnr4_ts.

Minimum battery voltage

Sample datalogger panel temperature

Average short-wave radiation (pyranometer up)

Average short-wave radiation (pyranometer down)

Average long-wave radiation (pyrge

Average CNR4 thermistor

Average CNR4 thermistor temperature

Average corrected long-wave radiation (pyrgeom

Average corrected long-wave radiation (pyrgeometer do

21