A.1.2 single-ended voltage measurement – Campbell Scientific AM25T 25-Channel Solid State Multiplexer User Manual

Page 26

background image

Appendix A. Single-Ended Measurements

03: BR Transform Rf[X/(1-X)] (P59)

;Calculate RTD resistance R/R0

1: 1

Reps

2: 1

Loc [ RefTemp_C ]

3: 10.025 Multiplier

(Rf)


04: Temperature RTD (P16)

;Calculate reference temperature

1: 1

Reps

2: 1

R/R0 Loc [ RefTemp_C ]

3: 1

Loc [ RefTemp_C ]

4: 1.0

Mult

5: 0.0

Offset


05: Beginning of Loop (P87)

;Loop through thermocouple channels

1: 0

Delay

2: 25

Loop

Count


06: Step Loop Index (P90)

;Set index to advance 2 locations/loop

1: 2

Step


07: Do (P86)

;Clock the AM25T twice

1: 41

Set Port 1 High

;Setting high then low is faster than pulsing w/21X


08: Do (P86)
1: 51

Set Port 1 Low


09: Do (P86)
1: 41

Set Port 1 High


10: Do (P86)
1: 51

Set Port 1 Low


11: Thermocouple Temp (SE) (P13)

;Measure the connected thermocouples

1: 2

Reps

2: 1

5 mV Slow Range

3: 1

SE

Channel

4: 1

Type

T

(Copper-Constantan)

5: 1

Ref Temp (Deg. C) Loc [ RefTemp_C ]

6: 2--

Loc [ TC_1 ]

7: 1.0

Mult

8: 0.0

Offset


12: End (P95)

;End of loop


13: Do (P86)

;Turn off AM25T

1: 52

Set Port 2 Low

A.1.2 Single-Ended Voltage Measurement

Connect two single-end sensors to the differential AM25T input channel.
Connect the sensor shields to the ground terminals next to the input channel. Up
to 50 single-ended voltage sensors may be measured by two single-ended
datalogger channels (Figure A-2).

A-4