AM25T Solid State Multiplexer

TABLE 6-3. RTD Excitation Voltage and Measurement Range for AM25T

Multiplexers Prior to Serial Number 1839

Datalogger

Temperature Range

Excitation voltage

Vs Range

Vx Range

21X

-40 to 85°C
-24 to 65°C

250 mV
350 mV

± 5 mV
± 5 mV

± 500 mV
± 500 mV

CR10(X)

-40 to 85°C
-11 to 52°C

350 mV
250 mV

± 7.5 mV
± 2.5 mV

± 2500 mV
± 250 mV

CR7

-40 to 85°C

350 mV

± 15 mV

± 500 mV

6.2.1 Example CR10(X) Programs

The two example CR10(X) programs in this section cover measuring
thermocouples and low level voltage sensors with the differential measurement
instruction. The 21X and CR7 are programmed similarly.

Input locations within the measurement loops must be indexed (--), see
Instruction 87 in the datalogger manual. To index a location, using Edlog ver
6.0 or greater, move the cursor to the Input location label and press “F4”. To
index a location with the datalogger keyboard, press the “C” key when
entering the location number and before the “A” key is pressed.

TABLE 6-4. Wiring for CR10(X) Examples

Function AM25T

CR10(X)

+12V Power

12 V

12 V

Power and Shield Ground

G

Clock CLK

C1

Reset RES

C2

RTD Excitation

EX

E1

Analog Ground

AG

AG

Common High

HI

1H

Common Low

LO

1L

Example 3. CR10(X) Program for Measuring 25 Type T Thermocouples or Voltage Sensors
Using a Differential Instruction

*Table 1 Program
01: 10

Execution

Interval

(seconds)


01: Set Port(s) (P20)

;Configure Control Ports for 1 millisecond pulse

1: 9999

C8..C5

=

nc/nc/nc/nc

2: 9933

C4..C1

=

nc/nc/1ms/1ms


02: Do (P86)

;Turn On AM25T

1: 42

Set Port 2 High

14