Campbell Scientific AM25T 25-Channel Solid State Multiplexer User Manual
Page 19

AM25T Solid State Multiplexer
03: Full Bridge (P6)
;Measure the output of the reference temp. full bridge
1: 1
Reps
2: 23
25 mV 60 Hz Rejection Range ;See Table 6-1 for older AM25T multiplexers
3: 1
DIFF
Channel
4: 1
Excite all reps w/Exchan 1
5: 1200
mV
Excitation
;See Table 6-1 for older AM25T multiplexers
6: 1
Loc [ RefTemp_C ]
7: -0.001 Mult
8: 0.09707 Offset
04: BR Transform Rf[X/(1-X)] (P59)
;Calculate RTD resistance R/R0
1: 1
Reps
2: 1
Loc [ RefTemp_C ]
3: 10.025 Multiplier
(Rf)
05: Temperature RTD (P16)
;Calculate reference temperature
1: 1
Reps
2: 1
R/R0 Loc [ RefTemp_C ]
3: 1
Loc [ RefTemp_C ]
4: 1.0
Mult
5: 0.0
Offset
06: Beginning of Loop (P87)
;Loop through thermocouple channels
1: 0
Delay
2: 25
Loop
Count
07: Do (P86)
;Clock the AM25T twice
1: 71
Pulse
Port
1
08: Do (P86)
1: 71
Pulse
Port
1
09: Thermocouple Temp (DIFF) (P14)
;Measure the connected thermocouple
1: 1
Reps
2: 21
2.5 mV 60 Hz Rejection Range
3: 1
DIFF
Channel
4: 1
Type
T
(Copper-Constantan)
5: 1
Ref Temp (Deg. C) Loc [ RefTemp_C ]
6: 2--
Loc [ TC_1 ]
7: 1.0
Mult
8: 0.0
Offset
10: End (P95)
;End loop
11: Do (P86)
;Turn off AM25T
1: 52
Set Port 2 Low
15