229 Heat Dissipation Matric Water Potential Sensor

0

100

200

300

400

500

200

100

0

100

200

10 degrees C
16 degrees C
18 degrees C
22 degrees C
24 degrees C
30 degrees C

matric potential (-kPa)

error (-kPa)

FIGURE 6-3. Measurement error for range of soil temperatures and wetter range of matric potential.

A temperature correction for the difference in temperature at time of calibration
and time of measurement is provided in the work of Flint et al., 2002. To
implement the correction, normalized temperature must be used for the
calibration variable. Normalized temperature is as defined in equation [3].

The correction procedure is an iterative method. Examples of implementing
the iterative routine with dataloggers are given in programming examples 5 and
6. The general sequence for using the temperature correction method is:

1. Determine

dry

T

Δ

and

wet

T

Δ

prior to calibration and with sensor

conditions as described in 6.2.1. The recommended

T

Δ

is the sensor

temperature after 30 seconds of heating minus the sensor temperature
after 1 second of heating.