229 Heat Dissipation Matric Water Potential Sensor

5.6 Example #2 — CR1000 with AM16/32-series Multiplexer,

CE4 and Sixteen 229 Sensors with Temperature
Correction

Table 5-2 shows wiring information for connecting multiple 229 sensors and
CE4 excitation module to an AM16/32 multiplexer and CR1000 datalogger.
See Figure 6-4 for a schematic of this wiring configuration.

TABLE 5-2. 229 Sensor and CE4 Wiring with CR1000 and AM16/32-series


229


107


Function


CR1000


CE4

Multiplexer
(4x16 mode)

Blue

229 Thermocouple High

ODD H
(1H, 3H, etc)

Red

229 Thermocouple Low

ODD L
(1L, 3L, etc)

Green

229 Heater High

EVEN H
(2H, 4H, etc)

Black

229 Heater Low

EVEN L
(2L, 4L, etc)

Clear

229

Shield

G

CE4

Power

12V

+12V

CE4

Ground

G

CE4

Enable

C3 CTRL

1H

COM

ODD

H

1L

COM ODD L

CE4 current excitation channel

Channel 1 COM EVEN H

CE4

Ground

COM EVEN L

AM16

Power

12V

12V

AM16/32

Ground

G

GND

AM16/32

Enable

C1

RES

AM16/32

Advance

C2

CLK

Red

107 Signal

SE 3

Black

107

Excitation

EX1

Purple

107

Signal

Ground

Clear

107

Shield

G