Campbell Scientific 229 Water Matric Potential Sensor and CE4/CE8 User Manual
Page 18
229 Heat Dissipation Matric Water Potential Sensor
;{CR10X}
;Program to read 1 229-L sensor
;Reading 1 sensor takes 30 seconds
*Table 1 Program
01: 60
Execution
Interval
(seconds)
1: If time is (P92)
1: 0
Minutes (Seconds --) into a
2: 60
Interval (same units as above)
3: 11
Set
Flag
1
High
2: If Flag/Port (P91)
1: 11
Do if Flag 1 is High
2: 30
Then
Do
3: Temp (107) (P11)
;Measure reference temperature
1: 1
Reps
2: 3
SE
Channel
3: 3
Excite
all
reps
w/E3
4: 1
Loc [ Ref_Temp ]
5: 1.0
Multiplier
6: 0.0
Offset
4: Thermocouple Temp (DIFF) (P14)
;Measure initial sensor temperature
1: 1
Reps
2: 1
2.5
mV
Slow
Range
3: 1
DIFF
Channel
4: 1
Type
T
(Copper-Constantan)
5: 1
Ref Temp (Deg. C) Loc [ Ref_Temp ]
6: 2
Loc [ Tinit_1 ]
7: 1.0
Mult
8: 0.0
Offset
5: Do (P86)
;Turn on CE8
1: 41
Set
Port
1
High
6: Excitation with Delay (P22)
;Wait one second before taking first reading
1: 1
Ex
Channel
2: 0000
Delay W/Ex (0.01 sec units)
3: 100
Delay After Ex (0.01 sec units)
4: 0
mV
Excitation
7: Thermocouple Temp (DIFF) (P14)
;Take 1 second temperature reading
1: 1
Reps
2: 1
2.5
mV
Slow
Range
3: 4
DIFF
Channel
4: 1
Type
T
(Copper-Constantan)
5: 1
Ref Temp (Deg. C) Loc [ ref_temp ]
6: 3
Loc [ T1sec_1 ]
7: 1.0
Mult
8: 0.0
Offset
14