229 Heat Dissipation Matric Water Potential Sensor

'CR1000
SequentialMode

Const Num229 = 4

'Enter number of 229 sensors to measure

Dim LoopCount
Public RefTemp_C, StartTemp_C(Num229), Temp_1sec_C(Num229)
Public Temp_30sec_C(Num229), DeltaT_C(Num229)
Public Flag(1) as Boolean

Units StartTemp_C()=Deg C
Units DeltaT_C()=Deg C

DataTable(Matric,Flag(1),-1)
Sample(Num229,StartTemp_C(),FP2)
Sample(Num229,DeltaT_C(),FP2)
EndTable

BeginProg
Scan(30,Sec,1,0)
PanelTemp

(RefTemp_C,250)

If IfTime (0,240,Min) Then Flag(1)=True 'Every 4 hours set Flag(1) high
If Flag(1) = True Then

'Flag(1) true triggers 229 readings

'Measure

starting

temperature

before

heating

TCDiff(StartTemp_C(),Num229,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0)
PortSet

(1,1

)

'Set C1 high to activate CE4

Delay

(0,1,Sec)

'Wait 1 second

'Measure

temperature

after

1

second

of

heating

TCDiff(Temp_1sec_C(),Num229,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0)
'Measure temperature after 30 second of heating
Delay (0,29,Sec)

'Wait 29 seconds more for total of 30 seconds heating

TCDiff(Temp_30sec_C(),Num229,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0)
PortSet

(1,0

)

'Set C1 low to deactivate CE4

For

LoopCount=1

to

Num229

'Calculate

temperature

rise

DeltaT_C(LoopCount)=Temp_30sec_C(LoopCount)-Temp_1sec_C(LoopCount)
'LoopCount=LoopCount+1
Next

LoopCount

EndIf

'Ends Flag(1) true condition

CallTable(Matric)

'Call Data Tables and Store Data

Flag(1)=False

'Set Flag 1 false to disable 229 measurements

NextScan
EndProg