beautypg.com

7 example #3 — cr10x with 229 sensor – Campbell Scientific 229 Water Matric Potential Sensor and CE4/CE8 User Manual

Page 17

background image

229 Heat Dissipation Matric Water Potential Sensor

'Measure temperature after 30 second of heating
TCDiff(Temp_30sec_C(i),1,mV2_5C,1,TypeT,RefTemp_C,True,0,_60Hz,1,0)

PortSet (3,0 )

'Set C3 low to deactivate CE4

DeltaT_C(i)=Temp_30sec_C(i)-Temp_1sec_C(i)'Calculate temperature rise
Call TempCorr

'Call temperature correction subroutine

dTcorr(i)=DeltaTcorr

Next i
EndIf

'Ends Flag(1) high condition

PortSet(1,0)

'Turn multiplexer Off

CallTable(Matric)

'Call Data Tables and Store Data

Flag(1)=False

'Disable 229 measurements

NextScan
EndProg

5.7 Example #3 — CR10X with 229 Sensor

Table 5-3 shows wiring information for reading a single sensor with a CR10X
datalogger, CE4 current excitation module, and CR10XTCR thermocouple
reference temperature sensor.

TABLE 5-3. 229 Sensor and CE4/CE8 Wiring with CR10XTCR

229

Function

CR10(X) CE4

CR10XTCR

Blue

229 Thermocouple High

1H

Red

229 Thermocouple Low

1L

Green

229 Heater High

Channel 1

Black 229

Heater

Low

Clear 229

Shield

G

CE4/CE8

Power

12V

+12V

CE4/CE8

Ground

G

CE4/CE8

Enable

C1

CTRL

CR10XTCR

Signal

2H

Red

CR10XTCR Signal Return AG

Clear

CR10XTCR

Excitation

E3

Black

13