beautypg.com

0 basic calibration – Red Lion ITMA User Manual

Page 6

background image

6

1 2 3 4 5 6 7

9

8

10

ON

1 2 3 4 5 6 7

9

8

10

ON

1 2 3 4 5 6 7

9

8

10

ON

1 2 3 4 5 6 7

9

8

10

ON

1 2 3 4 5 6 7

9

8

10

ON

2

1

ON

4

3

5 6 7 8 9 10

ON

3

1 2

5

4

6 7 8 9 10

12 to 42 VDC

Power Supply

±

4 to 20 mA

A

Indicator

Current Display

POWER/

OUTPUT

M2068B

+

4

5

-

6

Thermometer

+

TC

-

1 2

3

Source

+ -

Precision

Millivolt

Adjustable

ITMA

Basic Calibration Wiring

Step 4.1

Steps 4.3 to 4.15

Step 4.16

Step 4.17

Step 4.18

4.0 Basic Calibration

The Basic Calibration should only be performed with an ambient temperature

between 21°C and 29°C. The Basic Calibration was performed on the unit at the

factory and generally does not need to be done again. This procedure initializes the

unit by calibrating the input, and the Ice Point Compensation. The Basic Calibration

should be performed only if a condition exists as described in the “Calibration

Malfunction” section. After completion of this calibration, the unit needs to be scaled

in Field Calibration. The Basic Calibration procedure is described below.
Note: To abort this calibration and reset to the previous settings, set the BASIC CAL

switch OFF prior to the final setting of the OUTPUT CAL switch (Step 4.17) and

turn off power for 5 seconds. Then turn on power and the previous settings will be

loaded.

4.1 Connect a precision mV source with an accuracy of 0.02% to Terminal #1 (TC+

Input) and Terminal #2 (TC- Input). Set the ICE PT EN/DIS switch (#4), RANGE

(#9 ), TYPE (#6, #7, and #8), OUTPUT CAL (#1), and FIELD CAL (#2)

switches OFF. Set the BASIC CAL switch (#3) ON.

4.2 Apply power and allow a 30 minute warm-up period. [Current goes to 3.5 mA

(nominal)]

4.3 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.4 Input -9 mV and wait 5 seconds.

4.5 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.6 Input 6 mV and wait 5 seconds.

4.7 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.8 Input 22 mV and wait 5 seconds.

4.9 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.10 Input 41mV and wait 5 seconds.

4.11 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.12 Input 63 mV and wait 5 seconds.

4.13 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.14 Input 77 mV and wait 5 seconds.

4.15 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.16 Ice Point Calibration.

a. If ice point calibration is not desired, go to step 4.17.

b. To Enable ice point calibration, set the FIELD CAL switch (#2) ON.

1. Connect a precision thermometer (accuracy of 0.1°C) to the unused

terminal beside the TC Input terminals.

2. Allow 5 minutes for the temperature to equalize.

3. Using the temperature indicated by the precision thermometer, input an

equivalent 1 mV/°C signal to the TC Input terminals.

4.17 Set the OUTPUT CAL switch (#1) ON and then OFF.

4.18 Set the BASIC CAL switch (#3) and FIELD CAL switch (#2) OFF. [Current

increases to 3.6 mA (nominal) or more]

4.19 Perform a Field Calibration. (See Section 1.0)

TEST

OUTPUT CAL

FIELD CAL

BASIC CAL

ICE PT EN/DIS

OPEN SEN DN/UP

TC TYPE
TC TYPE
TC TYPE

RANGE
RANGE

DIP SWITCH SETTINGS:

ON = 1

OFF = 0

M2069A

1
2

4
5
6
7
8
9
10

ON

Test

Pad

RED LION CONTROLS

YORK, PA.

MADE IN U.S.A.

MODEL ITMA