Operating Environment
Operating temperature
-40 ... +60 °C (40 ... +140°F)
Pressure range
metal version
up to 200 bar
plastic version
up to 40 bar
Complies with EMC standard
EN61326-1: Industrial environment
Oil flow
some flow recommended
Inputs and Outputs
Minimum operating voltage
with current output
22…28 VDC
with voltage output
16…28 VDC
with RS-485
14…28 VDC
Supply current
normal measurement
20 mA + load current
External load for
current output
max. 500 Ω
voltage output
min. 10 kΩ
Analog outputs (2 standard)
current output
0 ... 20 mA, 4 ... 20 mA
voltage output
0 ... 5 V, 0 ... 10 V
Accuracy of analog outputs at 20 °C
± 0.05 % full scale
Temperature dependence of the analog outputs
current
± 0.005 %/°C full scale
voltage
Digital output
RS-485
Mechanics
Interface cable connector
M8 series 4-pin (male) with plug
(female) with straight or 90° angle
connector
Probe cable diameter
5.5 mm
Probe cable lengths
shielded cables
0.32 m, 3 m, 5 m and 10 m
90° angle cable
2 m and 5 m
LED cable
3 m
Housing material
metal
AISI 316L
plastic
PPS + 40% GF
Housing classification
IP 65 (NEMA 4)
Mechanical connections
Option 1
ISO G1/2"
Option 2
NPT 1/2"
Weight
Metal with ISO thread
200 g
Metal with NPT thread
200 g
Plastic with ISO thread
65 g