beautypg.com

Operating environment, Inputs and outputs, Mechanics – Vaisala MMT162 User Manual

Page 62

background image

User's Guide ______________________________________________________________________

60 __________________________________________________________________ M210934EN-C

Operating Environment

Operating temperature

-40 ... +60 °C (40 ... +140°F)

Pressure range

metal version

up to 200 bar

plastic version

up to 40 bar

Complies with EMC standard

EN61326-1: Industrial environment

Oil flow

some flow recommended

Inputs and Outputs

Minimum operating voltage

with current output

22…28 VDC

with voltage output

16…28 VDC

with RS-485

14…28 VDC

Supply current

normal measurement

20 mA + load current

External load for

current output

max. 500 Ω

voltage output

min. 10 kΩ

Analog outputs (2 standard)

current output

0 ... 20 mA, 4 ... 20 mA

voltage output

0 ... 5 V, 0 ... 10 V

Accuracy of analog outputs at 20 °C

± 0.05 % full scale

Temperature dependence of the analog outputs

current

± 0.005 %/°C full scale

voltage

Digital output

RS-485

Mechanics

Interface cable connector

M8 series 4-pin (male) with plug

(female) with straight or 90° angle

connector

Probe cable diameter

5.5 mm

Probe cable lengths

shielded cables

0.32 m, 3 m, 5 m and 10 m

90° angle cable

2 m and 5 m

LED cable

3 m

Housing material

metal

AISI 316L

plastic

PPS + 40% GF

Housing classification

IP 65 (NEMA 4)

Mechanical connections

Option 1

ISO G1/2"

Option 2

NPT 1/2"

Weight

Metal with ISO thread

200 g

Metal with NPT thread

200 g

Plastic with ISO thread

65 g