Chapter 6 ___________________________________________________ Calibration and Adjustment

VAISALA _______________________________________________________________________ 51

CHAPTER 6

CALIBRATION AND ADJUSTMENT

This chapter provides information that is needed in calibration and

adjustment of the product.

The MMT162 is fully calibrated and adjusted as shipped from factory.

The calibration interval depends on the application. It is recommended

that calibration is done always when there is a reason to believe that the

device is not within the accuracy specifications.

You can calibrate the MMT162 yourself or send it to Vaisala for

calibration. Calibration and adjustment is usually carried out through

serial line. One-point calibrations and adjustments can also be done by

using an MI70 indicator and Vaisala humidity calibrator HMK15.

Before the calibration, the used sensor needs to be cleaned with

instrument air to blow out existing oil or gently first flush with hepthane

(C

7

H

16

) and dry with instrument air to decrease response time, see

section Cleaning on page 47.

NOTE

It is important to clean the sensor before calibration as the oily

sensor can contaminate the salt bath and change the reference

condition.

Vaisala HUMICAP

®

hand-held moisture and temperature in oil meter

MM70 can also be used for calibration.