beautypg.com

Device information and status, Show device information, Show firmware version – Vaisala MMT162 User Manual

Page 29

background image

Chapter 4 ________________________________________________________________ Operation

VAISALA _______________________________________________________________________ 27

Device Information and Status

Show Device Information

The ? command outputs a listing of device information. If you wish to

display the listing for all devices on the current serial line, issue the

command ??.

?

??

Example:

?

MMT162 0.92

Serial number : G0000002

Batch number : D0720012

Module number : ????????

Sensor number : C6010000

Sensor model : Humicap L2

Cal. date : YYYYMMDD

Cal. info : NONE

Time : 00:21:19

Serial mode : STOP

Baud P D S : 4800 E 7 1

Output interval: 2 S

Serial delay : 0

Address : 0

Pressure : 1.0132 bar

Filter : 0.800

Ch1 output : 0 ... 20 mA

Ch2 output : 0 ... 20 mA

Ch0 error out : 0.000 mA

Ch1 error out : 0.000 mA

Ch0 aw lo : 0.00

Ch0 aw hi : 1.00

Ch1 T lo : -20.00 'C

Ch1 T hi : 80.00 'C

Show Firmware Version

Use the VERS command to display software version information.
Example:

vers

MMT162 1.02