Set voltage for led alarm, Extend analog output range – Vaisala MMT162 User Manual
Page 38

User's Guide ______________________________________________________________________
36 __________________________________________________________________ M210934EN-C
Hysteresis : 0.50
Level Lo : 0.20
Level Hi : 0.80
Channel 2:
Low Limit : OFF
High Limit : ON
Aerr : OFF
Quantity : aw
Limit Lo : 0.10
Limit Hi : 0.90
Hysteresis : 0.01
Level Lo : 0.20
Level Hi : 0.80
Set Voltage for LED Alarm
With the LED command, you can set a different voltage level for each
channel.
LED
Example:
led ?
Ch 1 Led Voltage : 2.80 V
Ch 2 Led Voltage : 2.80 V
Extend Analog Output Range
Use the AOVER command to allow the analog output channels to
exceed their specified range by 10 %. The scaling of the parameter
remains as before; the extra range is used for additional measurement
range in the wet end.
AOVER [ON/OFF]
Example:
aover on
: ON