Purpose procedure purpose procedure – Posey Bed 8070 User Manual
Page 30
30
Posey Bed 8070
Purpose
Procedure
Purpose
Procedure
To keep the Posey Bed clean and disinfected between complete bed and canopy cleanings.
Quick Clean the Posey Bed
machine Washing the Posey Bed
1. Wipe the fabric with cleaning agent
NOte: NeveR use a concentration
of cleaning agent greater than
recommended by the manufacturer.
1. Wash the Canopy on one gentle cycle in cold
water.
NOte: Water should be between 65-85°F
(18-30°C) for approximately 25 minutes.
2. Dry fabric.
3. Clean the Hook and Loop strips.
4. Clean the bed frame.
2. Dry fabric
3. Clean the Hook and Loop strips
4. Follow
transfer Patient into Bed
procedure.
a. Prepare a solution of diluted disinfectant or a
water-based germicide.
b. Use a damp cloth.
c. Rinse with water to remove excess cleaning
agent.
a. Use a mild detergent.
b. Load panels loosely into washing
machine.
c. Rinse thoroughly with water.
Hang dry at room temperature or
machine dry on air-only cycle (no heat).
Use a stiff brush to remove any lint or
foreign materials.
Wipe frame undercarriage with a damp
cloth.
NOte: Ensure cloth is damp, not wet.
Either wipe or air dry before use.
Use a stiff brush to remove any lint or foreign
materials.
aLWaYs wash the canopy between patients to help prevent transmission of infectious disease.
To keep the Posey Bed clean and disinfected between patients.
-
DO NOt use industrial extractor-type laundry equipment, as this may damage the zippers
and buckles.
-
NeveR use bleach, harsh cleaners or solvents.
-
aLWaYs attach hook and loop pieces together. Zip all zippers and connect Quick-Release
Buckles.
-
avOiD motor or bed controls when using water to clean bed frame, as this may damage the
electrical components.
-
uNPLuG the power cord before cleaning bed frame.
stePs
DO this
stePs
DO this