beautypg.com

Maintenance & storage – Sper Scientific 850038 Pure Water Meter User Manual

Page 37

background image

MAINTENANCE & STORAGE

35

a)Make sure the electrode is clean
Between measurements, rinse the
electrode with de-ionised water. If
the electrode has been exposed to a
solvent that is immiscible with water,
clean it with a solvent miscible with
water, e.g. Ethanol or acetone and
rinse carefully with water.

b)Store the conductivity cell carefully
Before storing the electrode, rinse it
carefully in de-ionised water.
Short-term storage: I

n de-ionised water.

Long-term storage: I

n de-ionised water

or store dry.

After long-term storage, leave the
electrode in de-ionised water for 8
hours before use.

c) Handle platinised cells carefully
Do not touch the black platinum layer
of the platinized cells.
To rinse the platinized cells,
dip the cell several times in a beaker
with demineralised water.

Be careful when rinsing platinized
cells with deionised water from a
water bottle. The force of the water can
remove some of the platinum layer
and consequently change the cell
constant.