beautypg.com

Sper Scientific 850038 Pure Water Meter User Manual

Page 31

background image

Select P5.0 and press ENTER to
advance to

P5.1. Press ENTER again

and the

cell constant value will flash.

Press UP/DOWN to switch the
value from 0.1,1.0,10.0, one by one.
The default value is 1.0. (Fig. Z)
Press ENTER to confirm the last
input and select P5.2 automatically.

Ranges available

K=0.1 K=1.0 K=10

0.00~19.99uS/0~9.99ppm

0.0~199.9uS/0~99.9ppm

0~1999uS/0~999ppm

0.00~19.99mS/0~9.99ppt

0.0~199.9mS/

0.0~99.9ppt

*

*

*

*

*

*

*

*

*

NOTE: The cell constant of the probe
sold with this meter is near K=1.0.

NOTE:
When using a cell constant K = 0.1,
the measured range will be only 1/10
of the range measured by K=1.
So, the lowest range will be 0~1.99us
(0~0.99ppm). Since only 5 ranges
are available, the highest range will
only be 0~19.99mS (0~9.99ppt).

.

Fig Z

K=

COND./TDS (Factor=0.5)

30

When using a cell constant K = 10,
the measured range will be 10 times
the range measured by K=1.
So, the highest range will be: 0 to
1999 mS (0.0 to 999 ppt). Since
only 5 ranges are available, the lowest
range will be 0 to 199.9uS (10.0 to
99.9 ppm)