uS

The LCD will display E03 if the
measured conductivity/TDS values
are beyond the limit value.
S

elect a range within limits.

The meter will reset to the Auto-
ranging function when it is turned off.
The

manual ranging function must be

set

each time the meter is turned on.

Type

P1.1

P1.2

P1.3

P1.4

P1.5

Range

Ist Range

2nd Range

3rd Range

4th Range

5th Range

Cell Constant
=0.1

Cell Constant
=1.0

Cell Constant
=10

0~1.99uS/ppm

0~19.99uS/ppm

0~199.9uS/ppm

0~1999uS/ppm

0~19.99mS/ppt

0~19.99uS/ppm

0~199.9uS/ppm

0~1999uS/ppm

0~19.99mS/ppt

0~199.9mS/ppt

0~199.9uS/ppm

0~1999uS/ppm

0~19.99mS/ppt

0~199.9mS/ppt

0~1999mS/ppt

The digits flash on the display.

Fig. P

CON

24

Press the UP or DOWN key to select
from P1.1 to P1.5 and then press
ENTER to save.
NOTE: Press ENTER to select
P1.1or other ranges, LCD will flash
the limit value (Fig. P). Press the
ENTER

key to confirm. The meter will

return to normal measure mode.